TPP0101 and TPP0201 Series 10X Passive Probes Information
1.
Connect the probe to the oscilloscope channel that you plan to use for your
measurements.
2.
Connect the probe to the probe compensation output terminals on the
oscilloscope front panel.
WARNING.
To avoid electric shock, only connect to the Probe Comp signal on the
oscilloscope when making this adjustment.
3.
Push
Autoset
or otherwise adjust your oscilloscope to display a stable
waveform.
4.
Adjust the trimmer in the probe until you see a perfectly
fl
at-top square wave
on the display. (See illustration.)
WARNING.
To avoid electric shock, only use the insulated adjustment tool when
making compensation adjustments.
Connecting the Probe to the Circuit
Use the standard accessories included with the probe to connect to your circuit.
WARNING.
To avoid electric shock when using the probe or accessories, keep
fi
ngers behind the
fi
nger guard of the probe body and accessories.
To reduce risk of shock, ensure the ground lead and ground spring are fully mated
before connecting the probe to the circuit under test.
30
TBS1000 Series Oscilloscopes Installation and Safety Manual