Appendix A: Specifications
A- 14
RFA300A Measurement Set 8VSB
Environmental Characteristics
Table A- 7: Environmental characteristics
Characteristic
Description
Cooling airflow
Intake is from the front and sides of the instrument. Exhaust is to the bottom and rear of the
instrument.
Required Clearance
2 in. (50 mm) air space adjacent to the bottom of the instrument is required.
Use Rating
Rated for indoor use only.
Atmospherics
Temperature:
Operating (meeting accuracy
specifications)
+5
°
C to 45
°
C (32
°
F to 113
°
F), 30
°
C/hr max gradient, non-condensing (derated 1
°
C or
1.8
°
F per 1,000 ft. or 305 m above 5,000 ft. or 1524 m altitude)
Operating (without damage)
+5
°
C to 50
°
C (32
°
F to 122
°
F), 30
°
C (54
°
F)/hr max gradient, non-condensing (derated
1
°
C per 1,000 ft. or 305 m above 5,000 ft or 1524 m altitude)
With a floppy diskette media in the floppy drive the operating specification lower limit is
10
°
C (50
°
F).
Non-operating
--20
°
C to 60
°
C (--4
°
F to 140
°
F), 30
°
C (54
°
F)/hr max gradient, non-condensing.
Humidity
Operating
20% to 80% relative humidity, non-condensing. Max wet bulb temperature: 31
°
C or 87.8
°
F
(at 50
°
C or 122
°
F, maximum relative humidity is ~50%).
Non-operating
8% to 80% relative humidity, non-condensing. Max wet bulb temperature: 40
°
C or 104
°
F
(at 50
°
C or 122
°
F, maximum relative humidity is ~50%).
Altitude
Operating
Up to 10,000 ft. (3,040m), (derated 1
°
C or 1.8
°
F per 1,000 ft. [305 m] above 5,000 ft.
[1524 m] altitude)
Non-Operating
40,000 ft (12,190 m)
Содержание RFA300A
Страница 1: ...User Manual RFA300A Measurement Set 8VSB 071 0697 01 www tektronix com...
Страница 4: ......
Страница 10: ...Table of Contents vi RFA300A Measurement Set 8VSB...
Страница 102: ...Appendix C System Recovery C 12 RFA300A Measurement Set 8VSB...
Страница 108: ...Glossary Glossary 4 RFA300A Measurement Set 8VSB...
Страница 112: ...Index Index 4 RFA300A Measurement Set 8VSB...