Installation
1- 20
RFA300A Measurement Set 8VSB
Incoming Inspection
Incoming inspection consists of verifying basic operation of the measurement
set. The Power-on diagnostics check basic functionality. These diagnostics run
every time you power on the instrument. If a failure is detected, the Power-on
diagnostics dialog box opens with the failed test indicated.
Run the extended diagnostics to test functionality in more detail. To run
extended diagnostics, do the following procedure:
1.
Go to the Tools menu and select Diagnostics.
2.
Click the Extended Diagnostics tab.
3.
Select All Modules and Tests and then select One Time.
4.
Click Run and observe that a failure does not occur.
If a failure does occur, contact Tektronix using the information provided in the
Preface
of this manual or click on Support located under the Help menu.
Perform the procedure on page 1--18 titled
Selecting Channel Frequency
and
verify that the RF input signal has sufficient power and is connected to the
selected RF input (1 or 2).
Open each measurement window and verify that activity is occurring. For
example, to check Frequency Response and Group Delay do the following
procedure:
1.
From the Home window, click on the Frequency Response and Group Delay
icon to open the measurement window.
2.
Verify that a signal is visible in each waveform graphic within a few
seconds.
3.
Verify that the readout for both measurements is updating.
Perform the same procedure for each measurement.
Extended Diagnostics
Signal Power
Measurement
Functionality
Содержание RFA300A
Страница 1: ...User Manual RFA300A Measurement Set 8VSB 071 0697 01 www tektronix com...
Страница 4: ......
Страница 10: ...Table of Contents vi RFA300A Measurement Set 8VSB...
Страница 102: ...Appendix C System Recovery C 12 RFA300A Measurement Set 8VSB...
Страница 108: ...Glossary Glossary 4 RFA300A Measurement Set 8VSB...
Страница 112: ...Index Index 4 RFA300A Measurement Set 8VSB...