Models 707B and 708B Switching Matrix Reference Manual
Section 7: TSP command reference
707B-901-01 Rev. B / January 2015
7-207
Details
This command sets the timer event detector to the undetected state and resets the overrun indicator.
Example
trigger.timer[1].clear()
Clears trigger timer 1.
Also see
(on page 7-207)
trigger.timer[N].count
This attribute sets the number of events to generate each time the timer generates a trigger event.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Recall setup
Trigger timer N reset
Create configuration script
1
Usage
count
= trigger.timer[
N
].count
trigger.timer[
N
].count =
count
count
Number of times to repeat the trigger (0 to 1,048,575)
N
Trigger timer number (1 to 4)
Details
If
count
is set to a number greater than 1, the timer automatically starts the next trigger timer delay
at the expiration of the previous delay.
Set
count
to zero (0) to cause the timer to generate trigger events indefinitely.
Example
print(trigger.timer[1].count)
Read trigger count for timer number 1.
Also see
(on page 7-206)
(on page 7-208)
(on page 7-211)