Performance Verification & Adjustment Procedures
Page 9 of 19
Procedure #: CP1005 Revision: A
6.
SINE WAVE DISTORTION
a.
Set-Up:
RANGE
100K
FUNCTION
Sine Wave
VOLTS OUT
Button Out
FREQ/10
Button Out
FREQ DIAL
1.0
b.
Connect MAIN OUTPUT to Distortion Analyzer via coax cable and 50 ohm
terminator.
c.
Set the Distortion Analyzer INPUT RANGE to Auto Range, and the FUNCTION
settings to Volts, Auto Range and THD+N.
d.
CHECK- For less than 1% distortion for the frequencies listed in Table 5:
Table 5
RANGE
FREQ DIAL
PER CENT
DISTORTION
100K
1.0
< 1 %
10K
1.8
< 1 %
1K
1.8
< 1 %
100
1.8
< 1 %
10
1.0
< 1 %
7.
FREQUENCY DIAL ACCURACY
a.
Set-Up:
AMPLITUDE
CW
RANGE
1K
FUNCTION
Triangle
VOLTS OUT
Button In
FREQ/10
Button Out
FREQ DIAL
3.0