Performance Verification & Adjustment Procedures
Page 10 of 19
Procedure #: CP1005 Revision: A
b.
Connect the CFG253 MAIN OUTPUT to the input of a Frequency Counter.
Check the frequency reading for each set-up listed in Table 6:
Table 6
RANGE FREQ DIAL
COUNTER READING
1 M
3.0
2.85 MHz - 3.15 MHz
1 M
0.3
150 KHz - 450 KHz *
100 K
3.0
285 KHz - 315 KHz
10 K
3.0
28.5KHz - 31.5KHz
1 K
3.0
2.85KHz - 3.15KHz
100
3.0
285 Hz - 315 Hz
10
3.0
28.5 Hz - 31.5 Hz
1
3.0
2.85 Hz - 3.15 Hz
*
The Frequency reading accuracy is 5% of full scale. 5% of 3 MHz is 150
KHz
c.
Disconnect cable from the Counter and CFG253.