CFG253
Performance Verification & Adjustment Procedures
Page 5 of 19
Procedure #: CP1005 Revision: A
l.
Set the FUNCTION to Square Wave. Use the CFG253 and test scope settings in
Table 1 to check the MAIN OUTPUT AMPLITUDE for the indicated settings :
TABLE 1
Function
Range
Volts Out
button
Freq.
Dial
Test Scope
Volts/Div
Test Scope
Display
Square Wave
1K
Out
3.0
2V/Div
> 5 Div
Sine Wave
1K
Out
3.0
2V/Div
> 5 Div
Sine Wave
1K
Out
0.3
2V/Div
> 5 Div
Square Wave
1K
Out
0.3
2V/Div
> 5 Div
Triangle
1K
Out
0.3
2V/Div
> 5 Div
m.
Remove the 50 ohm terminator and connect the coax cable directly to the test
scope input.
n.
Set the test scope Time/Div to 1uSec.
o.
Check the test scope display for each setting in Table 2:
TABLE 2
Function
Range
Volts
Out
Freq.
Dial
Volts/D
iv
Display
Triangle
1M
Out
3.0
5V/Div
> 4 Div
SquareWave
1M
Out
3.0
5V/Div
> 4 Div
Sine Wave
1M
Out
3.0
5V/Div
> 4 Div
p.
Press the VOLTS OUT button in. Check the test scope display for the settings in
Table 3:
TABLE 3
Function
Range
Volts
Out
Freq.
Dial
Volts/Div
Display
Sine Wave
1M
In
3.0
0.5V/Div
> 4 Div
Square Wave
1M
In
3.0
0.5V/Div
> 4 Div
Triangle
1M
In
3.0
0.5V/Div
> 4 Div