Maintenance
12
A6304XL Instructions
2. Place the probe on a flat surface with the screw heads facing up.
3. Using a
3
/
32
inch Allen wrench, remove the eight retaining screws from the
case.
4. While holding the squeeze handle and slide, lift the top part of the probe
body off.
5. Remove the slide.
WARNING.
There is a spring in the squeeze handle that can pop out and cause
personal injury unless care is exercised when disassembling the probe.
6. Lift the cable out of the probe handle and insert a pair of pliers in the handle
as shown in Figure 7.
Body Half
Squeeze
Handle
Torsion Spring
Figure 7: Removing the Handle
7. While maintaining a firm grip on the pliers, carefully lift the squeeze handle
and gear out of the probe. See Figures 7 and 9.
8. To remove the current transformer, lift the assembly out of the probe as
shown in Figure 8 and unplug it from the circuit board.