![Tektronix A6304XL Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/tektronix/a6304xl/a6304xl_manual_1078989017.webp)
Specifications
A6304XL Instructions
9
Table 2:
Mechanical Characteristics
Probe Weight
3.2 kg (7 lbs)
Cable Length
8 m (26 ft)
Jaw Dimensions
See Figure 6
Maximum Wire Size
21 mm (0.83 in)
Figure 6: A6304XL Probe Jaw Dimensions (Nominal)
Table 3: Environmental Characteristics
Operating Temperature
0
°
C to 50
°
C (32
°
F to 122
°
F)
Storage Temperature
–40
°
C to 75
°
C
(–40
°
F to 167
°
F)
Humidity
Operating
Storage
30
°
C to 50
°
C at 90 to 95% RH
30
°
C to 60
°
C at 90 to 95% RH
Altitude
Operating
Storage
2,000 m (6,416 ft)
15,380 m (50,000 ft)
Transportation
The packaged product qualifies under the
Distribution Cycle 1, Assurance Level II for
packaged products 0 – 20 lbs. Test 2 for
Warehouse and Vehicle Stacking
(Compression) is omitted.
Tektronix Design Standard 062-2858-00,
Rev. B, Section 17.
Mechanical Shock
500 g. Half sine. (200 g in axis towards handle
and nose) two axes. Three shocks on each
axes for 1 ms duration. Total of 18 shocks.
Random Vibration
Class 3: 3 Axes, 60 Min. Total, 20 Min/Axis
(10 Min. Operating, 10 Min. Nonoperating)