Fig. 6 -2 1 . Typical display for Sensitivity check (Step 1 3 ).
i.
When the 100 kHz RESOLUTION display meets the
sensitivity requirements given in Table 6-2, the 1 kHz RESOLU
TION display w ill also be w ithin its sensitivity requirements.
14. Adjust and Check Dynamic Ranges
O
a. Connect the equipment as shown in Fig. 6-20.
■b. Preset the controls of the Spectrum Analyzer as follows.-
Position the trace to the
bottom line o f the
graticule
All OFF
M H z/C M
1 (Outer Ring)
O ff (Down)
Fully counterclockwise
LIN
BAND 1
c. Set the generator for a 200 MHz signal.
d. Set the generator RF ATTEN control for a 6 centimeter
signal amplitude.
e. Decrease the generator output by 50 dB and turn the
Analyzer G A IN control fully clockwise. Check for a t least
6 centimeters of signal amplitude. Reduce the generator
output by 26 dB. The signal should still be visible.
f. Change the VERTICAL DISPLAY to LOG.
g. Set the generator RF ATTEN to — 100 dBm and adjust
the Analyzer G A IN control for 2 mm signal.
h. Set the generator RF ATTEN to —60 dBm and set the
Log Cal adjust R666 (see Fig. 6-22) for a 6 centimeter signal.
i. Return the generator RF Atten to — 100 dBm. The signal
should still be visable.
j. Change the VERTICAL DISPLAY to SQ LAW.
k. Set the generator RF Atten to —50 dBm.
POS
IF ATTEN
DISPERSION RANGE
DISPERSION-
COUPLED RESOLUTION
VIDEO FILTER
GAIN
VERTICAL DISPLAY
RF INPUT
Calibration— Type 1L20
Fig. 6 -2 2 . Location of Log Cal Adjustment R666 (Step 1 4 ) .
l
. Set the Analyzer G A IN control for a 6 centimeter dis
play.
m. Set the generator RF Atten to —63 dBm. The signal
should still be visible.
15. Check Display Flatness
a. Set up the equipment as shown in Fig. 6-20.
lb. Preset the Spectrum Analyzer controls as follows:
Position the trace to the
bottom line of the
graticule
Midrange (000)
POS
IF CENTER FREQ
IF ATTEN
DISPERSION RANGE
DISPERSION-
COUPLED RESOLUTION
VIDEO FILTER
VERTICAL DISPLAY
RF INPUT
RF CENTER FREQ
FINE FREQ
c. Set the generator for a
All OFF
M H z/C M
10
O ff (Down)
LIN
BAND 1
250 MHz
Fully Clockwise
—50 dBm 250 MHz signal.
d. Set the Analyzer G A IN control for a 5 centimeter signal
display.
e.
Turn the RF CENTER FREQ control from 225 MHz to
275 MHz and set at the frequency giving the highest am pli
tude signal.
f.
Set the Analyzer GAIN control for a 6 centimeter
signal display.
g. Set the 1 dB and 2 dB IF ATTEN switches to On and
note the display size. Return the IF ATTEN switches to OFF.
h. Tune the RF CENTER FREQ from 225 MHz to 275 MHz
and check that the signal display does not drop below the
size noted in step f.
<§)
6 -1 7
Содержание 1L20
Страница 4: ...Type 1L20 Fig 1 1 Type 1L20 Spectrum Analyzer...
Страница 16: ...NOTES...
Страница 18: ...Fig 3 1 Block Diagram of the Type 1L20 N Circuit Description Type 1L20...
Страница 32: ...Maintenance Type 1L20 4 10 Fig 4 6 Phase Lock and Recorder Detector Circuit Boards...
Страница 42: ...Calibration Type 1L20 4B 40 4D 4E 6 2 Fig 6 1 Recommended equipment for calibrating the Type 1L20...
Страница 43: ...Calibration Type 1L20 10 11 12 13 14 15 6 3 Fig 6 2 Recommended Calibration tools and equipment...
Страница 94: ......
Страница 97: ...N0I133S 3H ZHU i...
Страница 98: ...TYPE IL LO SPECTRUM ANALYZER A P H A S E LO C K U S C IR C U IT 2 PHASE LOCK CIRCUIT...
Страница 100: ...T 1 SW EEPER CIRCUITS...
Страница 102: ...N B IFA M PL 70 MHz OSC 6...
Страница 103: ...I o o z o o 5 1166 TYPE IL Z 0 IL 3 0 SPEC TR U M ANALYZER A VARIABLE RESOLUTION CIRCUITS T V A R IA B L E RESOL...
Страница 104: ...5 M H z AM PLIFIE R D E TECTO RS TYPE IL Z O IL 3 0 S P E C T R U M ANALYZER A OUTPUT AMPLIFIER OUTPUT AMPLIFIER...
Страница 105: ...TYPE 1L20 SPECTRUM ANALYZER...
Страница 106: ...FIG 2 REAR CHASSIS TYPE 1L20 SPECTRUM ANALYZER 1 F IG 2...
Страница 107: ...FIG 3 IF CHASSIS T i T L A TYPE 1L20 SPECTRUM ANALYZER F I G 3...
Страница 108: ...FIG 4 LOW PASS FILTER PHASE LOCK ASSEMBLY FIG 4...
Страница 110: ...OPTIONAL ACCESSORIES TYPE 1L20 SPECTRUM ANALYZER...
Страница 112: ...TYPE 1L20 TYPE 1L30 PARTS LIST CORRECTION CHANGE TO R823 30 0181 00 180 a 1W 10 Cl 366...