10. Display Flatness Check
a.
Preset the Spectrum Analyzer controls as follows:
POS
IF CENTER FREQ
IF ATTEN
DISPERSION RANGE
DISPERSION-
COUPLED RESOLUTION
VIDEO FILTER
VERTICAL DISPLAY
RF INPUT
RF CENTER FREQ
Position the trace to the
bottom line of the
graticule
Midrange (000)
All OFF
M H z/C M
5
O ff (down)
LIN
BAND 1
250 MHz
IF CENTER FREQ
DISPERSION RANGE
DISPERSION-
COUPLED RESOLUTION
VIDEO FILTER
VERTICAL DISPLAY
RF INPUT
Midrange (000)
M H z/C M
10
O ff (Down)
LOG
BANDS 2-5
b. Connect the Marker O utput connector o f the Type
184 Time-Mark Generator to the BANDS 2-5 RF INPUT con
nector of the Analyzer through a 10:1 50 0 Attenuator. (To
use a Type 180A, see step d below.)
c. Depress the Marker Selector switches for 10 ns (100 MHz)
and the modulation signal indicated in Table 5-4. (Be sure
the 10 ns switch stays depressed each time the modulation
signal is changed.)
b. A pply a —50 dBm 250 MHz signal from the UHF Signal
Generator to the BAND 1 RF INPUT connector.
c. Set the Analyzer G A IN control for a 5 centimeter signal
display.
d. Turn the RF CENTER FREQ control from 225 MHz to
275 MHz and set at the frequency giving the highest am pli
tude signal.
e. Set the Analyzer G AIN control for a 6 centimeter sig
nal display.
f. Set the 1 dB and 2 dB IF ATTEN switches to On and
note the display size. Return the IF ATTEN switch to OFF.
g. Tune the RF CENTER FREQ from 225 MHz to 275 MHz
and check that the signal display does not drop below the
size noted in step f.
h. A pply a —50 dBm 350 MHz signal from the signal gen
erator to the BANDS 2-5 RF INPUT connector and change
the RF INPUTS switch to BANDS 2-5.
i. Set the RF CENTER FREQ dial to 350 MHz.
j. Set the MIXER PEAKING CONTROL for maximum
signal amplitude.
k. Set the Analyzer G AIN control for a 5 centimeter
display.
l
. Turn the RF CENTER FREQ control from 300 MHz to
400 MHz and set at the frequency giving the highest am pli
tude signal.
m. Set the Analyzer G A IN control for a 6 centimeter signal
display.
n. Set the 1 dB and 2 dB IF ATTEN switches to On and
note the display size. Return the IF ATTEN switch to OFF.
o. Tune the RF CENTER FREQ from 300 MHz to 400 MHz
and check that the signal display does not drop below the
display size noted in step n.
11. Dispersion Accuracy Check
a. Preset the controls of the Spectrum Analyzer as follows:
POS
Position the trace to the
bottom line of the
graticule
d.
If a Type 180A Time-Mark Generator is to be used in
place of the Type 184, change the procedure of steps b
and c as follows.
1. Connect the Marker Out connector of the Type 180A
to the RF connector of the Harmonic M odulator Unit and
depress the 50 MC switch.
2. Connect a cable between the desired Modulation
signal banana jack and the Harmonic Modulator Modu 2
connector.
3. Connect the Harmonic Modulator Modu Harm Out
connector to the Analyzer BANDS 2-5 RF INPUT connector.
Fig. 5 -4 . Dispersion Accuracy displays.
5 -5
Содержание 1L20
Страница 4: ...Type 1L20 Fig 1 1 Type 1L20 Spectrum Analyzer...
Страница 16: ...NOTES...
Страница 18: ...Fig 3 1 Block Diagram of the Type 1L20 N Circuit Description Type 1L20...
Страница 32: ...Maintenance Type 1L20 4 10 Fig 4 6 Phase Lock and Recorder Detector Circuit Boards...
Страница 42: ...Calibration Type 1L20 4B 40 4D 4E 6 2 Fig 6 1 Recommended equipment for calibrating the Type 1L20...
Страница 43: ...Calibration Type 1L20 10 11 12 13 14 15 6 3 Fig 6 2 Recommended Calibration tools and equipment...
Страница 94: ......
Страница 97: ...N0I133S 3H ZHU i...
Страница 98: ...TYPE IL LO SPECTRUM ANALYZER A P H A S E LO C K U S C IR C U IT 2 PHASE LOCK CIRCUIT...
Страница 100: ...T 1 SW EEPER CIRCUITS...
Страница 102: ...N B IFA M PL 70 MHz OSC 6...
Страница 103: ...I o o z o o 5 1166 TYPE IL Z 0 IL 3 0 SPEC TR U M ANALYZER A VARIABLE RESOLUTION CIRCUITS T V A R IA B L E RESOL...
Страница 104: ...5 M H z AM PLIFIE R D E TECTO RS TYPE IL Z O IL 3 0 S P E C T R U M ANALYZER A OUTPUT AMPLIFIER OUTPUT AMPLIFIER...
Страница 105: ...TYPE 1L20 SPECTRUM ANALYZER...
Страница 106: ...FIG 2 REAR CHASSIS TYPE 1L20 SPECTRUM ANALYZER 1 F IG 2...
Страница 107: ...FIG 3 IF CHASSIS T i T L A TYPE 1L20 SPECTRUM ANALYZER F I G 3...
Страница 108: ...FIG 4 LOW PASS FILTER PHASE LOCK ASSEMBLY FIG 4...
Страница 110: ...OPTIONAL ACCESSORIES TYPE 1L20 SPECTRUM ANALYZER...
Страница 112: ...TYPE 1L20 TYPE 1L30 PARTS LIST CORRECTION CHANGE TO R823 30 0181 00 180 a 1W 10 Cl 366...