background image

Technology Solutions

T

 LAMP 1200A-100H

EK-

C

Instruction Manual

Handheld Ultrasonic Flow Meter

Document Number: IM-1200A-100H

www.tek-trol.com

Содержание TEK-CLAMP 1200A-100H

Страница 1: ...Technology Solutions T LAMP1200A 100H EK C Instruction Manual Handheld Ultrasonic Flow Meter Document Number IM 1200A 100H www tek trol com...

Страница 2: ...Customer Support 796 Tek Drive Crystal Lake IL 60014 USA Tel 1 847 857 6076 1 847 655 7428 www tek trol com No part of this publication may be copied or distributed transmitted transcribed stored in...

Страница 3: ...2 4 Dimensional Drawings 7 3 Installation 8 3 1 Correct Mounting Point 8 3 1 1 Clamp on Transducer 8 3 1 2 Clamp on mounting bracket transducer 9 3 2 Requirements for Measuring Flow meter Accuracy 9 3...

Страница 4: ...quence of such changes 1 2 2 Product Liability and Warranty The operator shall bear authority for the suitability of the device for the specific application The manufacturer accepts no liability for t...

Страница 5: ...which if not avoided may result in minor or moderate injury It may also be used to alert against unsafe practices NOTE Indicates that operating the hardware or software in this manner may damage it or...

Страница 6: ...typical commercial industrial environments The following considerations must be observed in selecting a location for the meter The ambient operating temperature range is 22 F 30 C to 176 F 80 C Do no...

Страница 7: ...of between 32 F 0 C to 320 F 160 C This flow meter can be virtually applied to a wide range of measurements A variety of liquid applications contain potable water cooling water river water raw sewage...

Страница 8: ...C Environment Humidity Convertor 85 RH Protection Class Flow Transducer IP67 Clamp On HS type for pipe size 4 12 5mm 100 mm HM type for pipe size 2 28 50mm 700 mm Other pipe size available on request...

Страница 9: ...200A 100H www tek trol com 7 2 4 Dimensional Drawings Fig 2 Top View of 1200A 100H Handheld Ultrasonic Flow Meter Fig 3 Front View of 1200A 100H Handheld Ultrasonic Flow Meter Fig 4 Bottom View of 120...

Страница 10: ...tallation of a flow Meter Installation at wrong point may influence measuring accuracy or flow meter service life or even damage the flow meter The axis for measuring electrode shall be approximate to...

Страница 11: ...ed with the medium Medium conductivity should be homogeneous to avoid severe disruption If chemical substances need be injected into the pipe the injection should be operated at upstream side of flow...

Страница 12: ...when the pipe diameter is above 8 200mm 3 3 3 W method Installation W method is usually used on plastic pipes with a diameter from 2 15mm to 50mm Downs tream transducer Tub e wal l Li nin g Upstr eam...

Страница 13: ...he gland nut on lead collar and take off blind secondly put gland nut and rubber ring on cable make the cable through lead collar finally straighten out cable screw gland nut tight to make rubber ring...

Страница 14: ...tically and horizontally while checking the varying signal strength stop at the highest position and then check the spacing of the transducers to make sure the spacing of the transducers 4 1 1 Signal...

Страница 15: ...t The user can choose only one type of output 4 3 1 The Connection of Digital Output Digital output has two connected points i e digital output connected point and digital ground point CAUTION Make su...

Страница 16: ...one of the new passwords that are provided by manufacturer Finally press the Compound Enter keys to work in Parameters Setting Way There are 6 Passwords in design and among them 4 for deferent operato...

Страница 17: ...window and press mode Use keys Up and down to scroll up and down to the intended pipe material and then press ENT 4 Press key down to enter M16 window press ENT key to enter the option selection ENT k...

Страница 18: ...ay today net totalizer Initial setup M10 Input outside perimeter of pipe M11 Input pipe outer diameter data range 0 18000mm M12 Input pipe wall thickness M13 Input pipe inner diameter M14 Choose the k...

Страница 19: ...ycle displaying method displaying the previous menu of 2 39 time interval is 8 seconds When there is no input operation it will automatically enter cycle displaying status Choosing Setup M40 Damper co...

Страница 20: ...p serial port parameter M63 Communication protocol choosing including compatible protocol choosing two options choosing MODBUS RTU means using binary system MODUS RTU protocol Choosing MODBUS ASCII pr...

Страница 21: ...tity unit 1 GJ default 2 KCal 3 KW 4 BTU imperial unit M85 Choose temperature signal origin if choosing inputting temperature signal by AI3 AI4 then need temperature transmitter that can output 4 20mA...

Страница 22: ...0 circuitry hardware parameter adjusting entrance only inputting password to enter following windows 1 4 20mA current loop calibration 2 AI3 inputting calibration of analogue input 4 mA 3 AI3 inputti...

Страница 23: ...circuit three connectors SIG 1 SIG 2 SGND of converter if no Empty Alarm is displayed then the converter works In this case it is possible that the conductivity of measured fluid may be limited or th...

Страница 24: ...r customers a comprehensive range of products and solutions for process power and oil gas industries Tek Trol provides process measurement and control products for Flow Level Temperature Pressure Meas...

Отзывы: