USE AND DESTINATION OF USE OF THE AUTOCLAVE
The autoclave is able to sterilize the three types of load provided for by the standard
EN13060:2009, especially:
AUTOCLAVE
18L
AUTOCLAVE
20L
AUTOCLAVE
24L
METAL OR SOLID MATERIALS
Instruments with no cavities and no
obstacles to the penetration of steam
max kg. 5
max kg. 6
max kg. 7
POROUS OBJECTS
Simple or composite materials that can
absorb fluids (fabrics, gowns, surgical
gauzes, dressings, etc)
max kg. 5
max kg. 6
max kg. 7
HOLLOW OBJECTS
Materials or devices with cavities,
obstructions, etc. These are subdivided
into two types, classified according to
the length and diameter of the cavity.
Approximately:
TYPE B: cannulas, tubes or devices
with large passages.
TYPE A: turbines, hand pieces and
devices with blind or small holes
max kg. 1,5
max kg. 1,75
max kg. 2
*Only for European Countries
The autoclave shall only have to be used for the sterilization of tools
and materials being compatible with the steam sterilization system.
Always make sure that the loads that need to undergo sterilization can
stand the temperatures scheduled for the selected cycle.
Содержание europa B pro
Страница 2: ...0434 Costruito da MEDILINE ITALIA s r l Via 8 marzo 4 43025 Corte Tegge Cavriago RE ITALY...
Страница 67: ...ENGLISH 25 Selecting a cycle from a list you will get a second mask containing all the details of the cycle...
Страница 107: ...FRAN AIS 25 En s lectionnant un cycle de la liste une deuxi me page s ouvrira indiquant tout le d tail du cycle...
Страница 187: ...ESPA OL 25 Seleccionando un ciclo se abre una segunda pagina con los detalles del ciclo mismo...
Страница 203: ...Fig A 07 01 04 03 02 00 05 06 08 09 10 11 12 13 14 15 16 17 18 19 20 21 22 SELECT CYCLE PUMP WATER B M1 B M2 B M3...
Страница 205: ......