ENGLISH
20
10.8 Discharge used water
When the used water level LED (Fig. A-pos.19) lights up, the used water collection tank
must be emptied.
If nothing is done, the operation of the autoclave is blocked.
Take the hose supplied (Fig.9), and insert it into the used water drain fitting located on the
front of the autoclave (Fig.A-pos.09). Put the other end of the hose in a container and
unscrew the ring nut, turning it anticlockwise, the water will fall into the container, emptying
the tank.
IMPORTANT:
A - The hose located in the receiving container, must never touch or be immersed in
the discharged water, otherwise sucking up will occur.
B – Always wait until the discharge water has been completely drained. The used water
maximum level LED turns off when some water still remains into the tank, consequently do
not considered it as a reference for this operation.
When drained, screw in the ring nut and remove the hose.
10.9
Interruption of a sterilization cycle
A sterilization cycle can be voluntarily interrupted by pressing the button
Start
for at least 2
seconds.
The autoclave will send out a beep and the display will show the alarm message A001.
For reset the alarm, press at the same time, the multifunction push buttons on the
Reset bar up to the cancelling of the same.
Содержание europa B pro
Страница 2: ...0434 Costruito da MEDILINE ITALIA s r l Via 8 marzo 4 43025 Corte Tegge Cavriago RE ITALY...
Страница 67: ...ENGLISH 25 Selecting a cycle from a list you will get a second mask containing all the details of the cycle...
Страница 107: ...FRAN AIS 25 En s lectionnant un cycle de la liste une deuxi me page s ouvrira indiquant tout le d tail du cycle...
Страница 187: ...ESPA OL 25 Seleccionando un ciclo se abre una segunda pagina con los detalles del ciclo mismo...
Страница 203: ...Fig A 07 01 04 03 02 00 05 06 08 09 10 11 12 13 14 15 16 17 18 19 20 21 22 SELECT CYCLE PUMP WATER B M1 B M2 B M3...
Страница 205: ......