ENGLISH
18
10.3 Characterisstics of the water to be used
TABLE SHOWING THE QUALITY LEVELS LAID BY THE UNI EN 13060: 2009
CEN STANDARD UNI EN 13060 : 2009
Evaporation residue
≤
10
mg/l
Silicon oxide
≤
1
mg/l
Iron
≤
0.2
mg/l
Cadmium
≤
0.005
mg/l
Lead
≤
0.05
mg/l
Remains of heavy metals apart from iron, cadmium,
lead
≤
0.1
mg/l
Chloride (Cl’)
≤
2
mg/l
Phosphate (P20s)
≤
0.5
mg/l
Conductivity (at 20°C)
≤
15
µ
s/cm
Ph value (acidity level)
5
÷÷÷÷
7,5
Appearance
Transparent, clear, without
deposits
Hardness (and alkaline earth ions)
≤
0.02
mmol/l
10.4 Loading of the materials in the autoclave
Arranging the materials to be sterilized on the provided trays, as follows:
••••
Do not superpose the materials
••••
Arrange the wrapped materials with the paper side facing upwards
••••
Never bring the materials into contact with the sterilization chamber or the autoclave
door
••••
Put scissors and dental forceps with blades open
After loading the materials, close the door of the autoclave. The display will show the icon
and the message DOOR CLOSED.
Содержание europa B pro
Страница 2: ...0434 Costruito da MEDILINE ITALIA s r l Via 8 marzo 4 43025 Corte Tegge Cavriago RE ITALY...
Страница 67: ...ENGLISH 25 Selecting a cycle from a list you will get a second mask containing all the details of the cycle...
Страница 107: ...FRAN AIS 25 En s lectionnant un cycle de la liste une deuxi me page s ouvrira indiquant tout le d tail du cycle...
Страница 187: ...ESPA OL 25 Seleccionando un ciclo se abre una segunda pagina con los detalles del ciclo mismo...
Страница 203: ...Fig A 07 01 04 03 02 00 05 06 08 09 10 11 12 13 14 15 16 17 18 19 20 21 22 SELECT CYCLE PUMP WATER B M1 B M2 B M3...
Страница 205: ......