DB1F020_rev8.docx
20
04
DATA AND TECHNICAL FEATURES
VERSION TECNO40
TECNO70
TECNO100-4L
EN ISO 10079 - 1
Surgical Suction Unit
HIGH VACUUM - HIGH FLOW
EN 60601 - 1
Protection against electric dangers
Class 1
Protection against direct/indirect contacts
TYPE B
Level of protection against the penetration
of substances
IPX1
Level of protection in zone AD-FT
( CEI64-4 )
NOT SUITABLE
Level of safety with anesthetic and /or
explosive mixtures
NOT SUITABLE
Use in field structures
NOT SUITABLE
Use on means of transport
NOT SUITABLE
Use for thoracic drainage
NOT SUITABLE
TECHNICAL FEATURES
Power supply
230V ~ 50 Hz
Absorbed electric power
240 W
420 W
405 W
Intensità di corrente
1.02 A
1.06 A
1.09 A
Intensità di corrente di spunto
3.09 A
5.07 A
5.08 A
Electric protection
2 fuses Ø 5x20
F 3.15 AL 250V
2 fuses Ø 5x20
F 6 AL 250V
ASPIRATION
Pump
WITH PISTON-NOT OILED
Pump nominal flow
50 Lt/min
90 Lt/min
120 Lt/min
Flux at use- max. established rate
40 Lt/min
70 Lt/min
100 Lt/min
Max. depression
-0.9 bar
POTS
N° pots
2
Capacity
Version 2 Lt and version 4Lt
Material POLYCARBONATE
OR
POLYSULFONE UPON REQUEST
Single use collection bag
UPON REQUEST
Non-return valve system
VALVE WITH FLOAT
HYDROFOBIC FILTER
Filtering Capacity
0.3 µm (99.999%)
Material
MICRO-GLASS FIBREPAPER
TUBE
Material,diameter
SILICON FOR MEDICAL USE ØINTERNAL= 8 ØEXTERNAL=
14
VACUUM REGULATION
Valve for vacuum regulation
PRESENT WITH SCALE
Vacuum-meter of graduated control
Ø 63 0 / -1 bar
DIMENSIONS
Length (mm)
540 mm
Depth (mm)
480 mm
Height with 1 l pot
850 mm
Weight
16 kg
19 kg
20 kg
ACOUSTIC LEVEL
/
LIMITS OF USE
SEE INSTRUCTION MANUAL
CONFORMITY
MDD 93/42 EEC - EN60601-1-2 (62-5) - EN ISO 10079-1
EXPECTED LIFE OF THE DEVICE
5 YEARS