DB1F020_rev8.docx
25
09
ANALYSIS AND RESEARCH OF FAILURES
N° FAULT
CAUSE
ASSISTANCE
1
IT DOESN’T
SUCK
THE POT IS BADLY
SCREWED DOWN
UNSCREW AND SCREW THE TOP, CHECK
GASKET’S INTEGRITY AND RIGHT POSITION
EXTERNAL PIPES WITH
FOLDING
REMOVE NARROWING ON THE PIPES
FILTER IS OBSTRUCTED
FILTER MUST BE REPLACED
2
THE FLOAT
DOESN’T
WRONG ASSEMBLING
CHECK THE CORRECT POSITION OF NON
RETURN VALVE
3 NOT
WORKING
EXTERNAL FUSE
INTERRUPTED
IN CASE OF NOT WORKING, ADDRESS TO
TECNO-GAZ
10
COMMERCIAL DATA, SPARE PARTS AND DOCUMENTATION
The positions refer to the electro-pneumatic outline (following chapter).
DESCRIPTION
REFERENCE
POWER SUPPLY CABLE
CE06006
KIT OF PIPES FOR EXTERNAL WIRING
SB1A138
HYDROFOBIC FILTER
CPF0029
POT WITH NON RETURN VALVE 2 L
CM84151
POT WITH NON RETURN VALVE 4 L
CM84152