background image

IST-2282.EO01.01

 

Istruzione / User’s Manual / Manuel d’utilisation 

Pag.5/9

 

 

 

 

TECNOCONTROL S.r.l. Via Miglioli 47 SEGRATE ( MI ) 

Tel. +39 02 26922890 - Fax +39 02 2133734

 

 

 

tion in Air”

 

then disconnect and connect the unit, finally try to replace a new 

“Cartridge Sensor

”. If the condition is 

not change, it will be necessary to replace the unit and/or send it back to the supplier to repair. 

Only  with  Red  set-up  “activate”  all  LED  activates  (

>24  mA  output  signal

):

 

this  happens  when  the  “

Cartridge 

Sensor

” is not working or gas concentration is out of scale (higher than 30%v/v O

2

) If there are not any gas leaks 

and the condition is not change, it will be necessary to replace the unit and/or send it back to the supplier to repair. 

INSTALLATION 

The detector must be accurately installed according to the national dispositions in force on the safety of the plants 
and installation of electric devices. 

Detector positioning

:

 The 

Fig.2

 

shows the instrument size. The unit must be positioned vertically with the sensor 

downwards. 

It should be fixed at 1.60 m from the floor (the O

2

 gas is lighter as air). 

Electrical Connection 

(

see Fig.2

)

: the maximum distance to install each detector from the Central Unit is show in 

the 

Table 1

. Normally use a tree wire shielded cable. 

The terminals 

(+ - S

), are plug-in type, it is necessary to extract them to make the connection. Pay attention when 

you insert them again, being polarized.  

Note: Dip-Switch should be set with instrument powered off. Dip-Switch 1 settled ON activates ALARM Led 

indication (see Table 2) (

see also Operational Description > Faults

). 

Important:

 

Once  installation  is  completed,

 

power  up  the  unit,  wait  about 

20÷30  minutes

  and  then  to  adjust  the 

sensor to the environment, 

only if it is necessary

,

 

carry out the 

Calibration in Air 

(

see

 Test and Calibration

).

 

COMPATIBILITY  WITH  OTHER  CENTRAL  UNITS:  In  case  of  a  central  unit  other  than  Tecnocontrol, 

please 

verify the max load resistor as shown in 

Fig.4

The max load resistor is 50 ohm with 12Vdc power supply, while is 

400 ohm with 24Vdc power supply. 

We suggest using 24Vdc power supply. 

WARNING 

Average life

: The sensitive element used in this detector has an excellent stability in time. In fresh air and in nor-

mal working condition the 

sensor's life is

 

about 2 years

. After this period the yellow LED “

FAULT

” flashes every 4 

seconds, is necessary replacing the “

Cartridge Sensor

”. 

Periodical  testing

:  we  advise  to  carry  out  working  tests  every  6  or  12  months  as  explained  on  page  4  chapter 

Tests and Calibration

”. 

Note:

 the detector is not able to detect Oxygen gas occurring outside  the room where it is installed, either inside 

walls or under the floor. 

Sensor  operating  limits: 

the  electro-chemical  cell,  in  any  case,  should  not  be  exposed  to  temperatures  lower 

than -40° or higher than 60°C. The electrochemical cell should not  be exposed  to an over-range concentration  of 
30 % O

2

. If this condition occurs, we suggest replacing the sensor. 

Safety:

 Oxygen sensors contain a small quantity potassium acetate solution which is corrosive. Normally this mate-

rial would only leak out as a result of either mechanical damage or an incorrect instrument use beyond the limits of 
its operational functioning. These sensors also contain small amounts of lead, lead oxide, platinum, silver, carbon 
and antimony. If the user comes into contact with the oxygen sensor contents the affected area should be washed 
with a copious supply of water. 

Cross  sensitivity  to  other  gases:

  The  electrochemical  cell  employed  has  a  good  resistance  towards  products 

such  as  sprays,  detergents,  ammonia,  glues  and  paints. 

However,  in  case  of  products  containing  substances  in 

great quantity, these could interfere with the Sensor and cause false alarms.

 We recommend ventilating the room 

when products like these are used. Volatile organic solvents in great quantity cause false alarms, and then the cell 
needs one or more hours in clean air to recover the normal sensitivity conditions. 

TEST and CALIBRATION 

PAY  ATTENTION

This  procedure  has  to  be  made  with  extreme  attention  and  by  authorized  and  trained 

people;  because  starting  this  procedure  it  will  increase  mA  Output  causing  the  activation  of  connected 
alarm devices to the gas Central Unit. 

Operation Check, Calibration in Air and Fine mA ADJ Output

:

 are different code protected functions. To access 

these functions is necessary to insert the relevant “

Code

" through the keys F1s and F2. To have the key pressure 

recognized, hold  pressing  it for around a second (

until the Green Led doesn't switch off for a moment

). Then the next key 

can be pressed. In case of error all it takes is waiting around 10 seconds and the sequence is automatically erased. 
The 

Calibration Check

 is a function code free. 

Calibration Kit, Sample Gas Bottles

 (

for Calibration Check

) please, using a mixture: 

about 16%v/v Oxygen (O

2

) in Nitrogen

 

It is possible to use either the disposable cylinders with adjust valve or the high pressure ones with reduction gear. 

It is necessary to use 

the calibration kit 

Tecnocontrol model TC011

 

(

for non-corrosive gases

or TC014 (

Stainless steel

). 

“INSTRUMENT OPERATION CHECK” (

Check Code

: F2, F2, F1, F1)

this function allows to effect a functional test 

of the equipment. After having put the system in safety and inserted the "

Code Test

", all Led are switched off. Then 

they will switch on in sequence, the Led, from the yellow up to the red. At the end all the Led will remain lighted for 
around 5 seconds, then the central returns at the conditions of normal operation. It is advisable to perform this op-
eration every 6-12 months according to the use. 

Note: this function is not working if the red Led have already turned on.

 

“CALIBRATION IN AIR” (

Calibration Code

: F2, F2, F2, F1, F2, F1)

this function allows completely recalibrate the 

sensor at 20.9%v/v in air. This procedure when it is launched will automatically until the end. 

Содержание TS282EO

Страница 1: ... 1 del segnale signal Precisione Accuracy Precision 2 del segnale in aria signal in air du signal en air Deriva a lungo termine Long time drift Dérive à long terme 2 segnale mese signal per month du signal chaque mois 5 sulla vita operativa over Operating Life sur sa vie Temp umidità di funzionamento Operation Temp Humidity Température et hygrométrie de fonctionnement 20 50 C 15 90 RH non condensa...

Страница 2: ...ntrazione di Gas O2 supera il 22 5 oppure scende sotto il 19 5 solo se è stato abilitato con il Dip Switch n 1 posizionato su ON I Guasti possibili sotto elencati sono indicati dall accensione del Led giallo FAULT e portando l uscita S a 0mA Il Led Giallo si accende ogni 4 secondi con il Led Verde acceso per avvisare che la Cartuccia Sensore ha su perato il suo limite teorico di vita e non è più g...

Страница 3: ...o premuto per circa un secondo finche non si spegne per un attimo il Led Verde Dopodiché si può passare al pulsante successivo In caso d errore basta aspettare circa 10 secondi e la sequenza è automaticamen te cancellata Kit di Taratura e Bombole con Miscela Gas La miscela da utilizzare è Gas Ossigeno O2 a circa 15 v v in azoto È possibile usare bombole monouso con valvola d erogazione sia le rica...

Страница 4: ...TED GAS Oxygen O2 is a colourless odourless gas a little heavier than air its density to air is 1 1 Oxygen is not a flam mable gas it is a strong oxidant which reacts with reducing and flammable materials increasing the risk of fire and explosion We all need to breathe the oxygen in air to live Air is made up of several different gases including oxy gen Normal ambient air contains an oxygen concen...

Страница 5: ...sed to an over range concentration of 30 O2 If this condition occurs we suggest replacing the sensor Safety Oxygen sensors contain a small quantity potassium acetate solution which is corrosive Normally this mate rial would only leak out as a result of either mechanical damage or an incorrect instrument use beyond the limits of its operational functioning These sensors also contain small amounts o...

Страница 6: ...A corresponding to 0 5 O2 but it should be used only at new installation and if after the Calibration in Air clean air the control unit connected to TS282EO indicates a value other than 20 9 O2 After entering the ADJ Code the yellow LED starts to flash the output value increases pressing the F1 key and decreases pressing F2 Hold down the two F1 and F2 keys simultaneously for 2 seconds to confirm a...

Страница 7: ... nationales en vigueur pour les installations électriques et les normes de sûreté des installations Montage en Fig 1 sont indiquées les dimensions Le détecteur doit être monté en position verticale avec le cap teur tourné vers le bas Positionnement du TS282EO il doit être fixé à environ 1 60 m du plancher l oxygène ayant une densité proche de celle de l air Liaisons électriques Fig 2 La distance m...

Страница 8: ...eil portant la sortie S au 0mA L étalonnage doit être exécuté en air propre exclusivement milieu sans la présence de gaz toxiques ou autres A l aide des touches exécuter le Code d Etalonnage Attendre que les LED Jaune et Verte s allument et la Led Rouge clignote La LED Rouge reste ensuite en feu fixe pour indiquer que la procédure est en fonction Attendre que la LED rouge s éteigne au moins durant...

Страница 9: ...t 0 75 mm2 26 Ω km 100 m 300 m 1 mm2 20 Ω km 150 m 400 m 1 5 mm2 14 Ω km 200 m 500 m 2 5 mm2 8 Ω km 400 m 800 m Tabella 2 Table 2 Tableau 2 S1 SET Dip Switch Led Allarme Alarm Led Led d alarme 1 2 3 4 ON ON ON ON OFF ON OFF OFF OFF ON Fig 1 Dimensioni Size Dimensions 1 2 3 4 ON OFF Dip Switch SET Fig 2 Schema di collegamento Wiring diagram Schéma TESTmV TP2 TP1 S 12 24V mA Centrale rivelazione gas...

Отзывы: