background image

 

TECNOLOGIC spa - TLK 43

 

- OPERATING INSTRUCTIONS - Vr.02 - ISTR 06267 - PAG. 5

 

 
Tecnologic S.p.A. and its legal representatives do not assume 
any responsibility for any damage to people, things or animals 
deriving from violation, wrong or improper use or in any case 
not in compliance with the instrument’s features. 
 
3.4 - ELECTRICAL WIRING DIAGRAM 

 
 
 
 

4 - FUNCTIONS

 

 

4.1 – MEASURING AND VISUALIZATION 

All the parameters referring measurements are contained in the 
group 

]

InP”. 

By using par. 

“HCFG”,

 it is possible to select the input signal type 

which may come: from a thermocouple (tc), a thermo-resistance or 
a thermistor (rtd), from a transducer with normalised analogue 
signal in current (I) or tension (UoLt) or also from a signal coming 
from the communication serial line of the instrument (SEr).  
Once the signal type has been selected, it is necessary to set the 
type of input probe on par. 

“SEnS”,

 which can be : 

- for thermocouples J (J), K (CrAL), S (S), B (b), C (C), E (E), L (L), 
N (n), R (r), T (t) or for infrared sensors serie TECNOLOGIC IRS – A 
range - with linearization J (Ir.J) or K (Ir.CA) 
- for thermoresistances Pt100 IEC (Pt1) or thermistors PTC KTY81-
121 (Ptc) or NTC 103AT-2 (ntc) 
- for normalised signals in current 0..20 mA (0.20) or 4..20 mA 
(4.20) 
- for normalised signals in tension 0..50 mV (0.50), 0..60 mV (0.60), 
12..60 mV (12.60), 0..5 V (0.5), 1..5 V (1.5), 0..10 V (0.10) or 2..10 V 
(2.10). 
We recommend that the instrument be switched on and off 
whenever these parameters are modified, in order to obtain a 
correct measurement. 
For the instruments with input for temperature probes (tc, rtd) it is 
possible to select the unit of measurement  (°C, °F) through par. 

“Unit”,

 and the desired resolution (0=1°; 1=0,1°) through par. 

“dP”

Instead, with regards to the instruments with normalised analogue 
input signals, it is first necessary to program the desired resolution 
on par. 

“dP”

  (0=1; 1=0,1; 2=0,01; 3=0,001) and then, on par. 

"SSC"

, the value that the instrument must visualise at the beginning 

of the scale (0/4 mA, 0/12 mV, 0/1 V o 0/2 V) and, on par. 

"FSC", 

the value that the instrument must visualise at the end of the scale 
(20 mA, 50 mV, 60 mV, 5 V or 10 V). 
In the case of infrared sensors (TECNOLOGIC IRS-"A" range), by 
programming the sensor as "Ir.J" or "Ir.CA", the par.  "

rEFL

" is also 

present and it allows the correction of possible measuring errors 
caused by the environment lighting and by the reflectivity of the 
material. This parameter should be programmed with a high value if 
the material to be measured is particularly bright / reflective and 
must be reduced if the surface is particularly dark / not reflective, 

keeping in mind however that for most materials, the recommended 
value is within 1.00 and 0.80.   
The instrument allows for measuring calibration, which may be used 
to recalibrate the instrument according to application needs, by 
using par. 

“OFSt”

 and 

“rot”.

 

Programming par. “rot”=1,000, in par. “OFSt” it is possible to set a 
positive or negative offset that is simply added to the value read by 
the probe before visualisation, which remains constant for all the 
measurements. 
If instead, it is desired that the offset set should not be constant for 
all the measurements, it is possible to operate the calibration on  
any two points.  
In this case, in order to decide which values to program on par. 
“OFSt” and “rot”, the following formulae must be applied : 

“rot” = (D2-D1) / (M2-M1) “OFSt” = D2 - (“rot” x M2)

 

where: 
M1 =measured value 1 
D1 = visualisation value when the instrument measures M1 
M2 =measured value 2 
D2 = visualisation value when the instrument measures M2 
It then follows that the instrument will visualise : 

DV = MV x “rot” + “OFSt”

 

where:  DV = visualised value  

MV= measured value 

Example 1: It is desired that the instrument visualises the value 
effectively measured at 20° but that, at 200°, it visualises a value 
lower than 10° (190°).     
Therefore :  M1=20 ;  D1=20 ;  M2=200 ;  D2=190 
“rot” = (190 - 20) / (200 - 20) = 0,944 
“OFSt” = 190 - (0,944 x 200) = 1,2

 

Example 2: It is desired that the instrument visualises 10° whilst the 
value actually measured is 0°, but, at 500° it visualises a 50° higher 
value (550°). 
Therefore :  M1=0 ;  D1=10 ;  M2=500 ;  D2=550 
“rot” = (550 - 10) / (500 - 0) = 1,08 
“OFSt” = 550 - (1,08 x 500) = 10

 

By using par.

 “FiL” 

it is possible to

 

program time constant of the  

software filter for the input value measured, in order to reduce noise 
sensitivity (increasing the time of reading). 
In case of measurement error, the instrument supplies the power as 
programmed on par.  

“OPE”. 

This power will be calculated according to cycle time programmed 
for the PID controller, while for the ON/OFF controllers the cycle

 

time is automatically considered to be equal to 20 sec. (e.g. In the 
event of probe error with ON/OFF control and “OPE”=50, the control 
output will be activated for 10 sec., then it will be deactivated for 10 
sec. and so on until the measurement error remains.). 
By using par. 

“InE”

 it is also possible to decide the conditions of the 

input error, allowing the instrument to give the power  programmed 
on par. “OPE” as output. 
The possibilities of par. “InE” are : 
= Or : the condition occurs in case of over-range or probe breakage 
= Ur : the condition occurs in case of under-range or probe 
breakage 
= Our : the condition occurs in case of over-range or under-range or 
probe breakage 
Using par. 

“diSP”, 

located in the group 

]

PAn”, 

it is possible to set 

normal visualization of the SV display which can be the active Set 
Point (SP.F), the control power (Pou), the Set Point operating when 
there are active ramps (SP.o) or alarm threshold AL1, 2 or 3 (AL1, 
AL2 or AL3).

 

 
4.2 - OUTPUT CONFIGURATION 

The instrument’s outputs can be programmed by entering the group 
of parameters 

]

O1”, “

]

O2”, “

]

O3”, “

]

O4”

, where different 

parameters (depending on the type of outputs –digital or analogue- 
available on the instrument) are located.

 

Note:

 In the following examples, the number of outputs is 

generically indicated  with 

n

 

- DIGITAL OUTPUTS relay or SSR type : 

Within the selected group only the par. 

“OnF” 

will be present

This parameter can be set for the following functions : 
= 1.rEG : Main control output 

Содержание TLK 43

Страница 1: ...also instructions for its maintenance and use we therefore recommend that the utmost attention is paid to the following instructions Though this manual has been issued with the greatest care TECNOLOG...

Страница 2: ...ualized on 4 green displays while the outputs status is indicated by 4 leds The instrument provides for the storage of 4 Set Points and can have up to 4 outputs relay type or can drive solid state rel...

Страница 3: ...e visualised on the SV display by pushing the P key Again by using the UP and DOWN keys it is possible to select the desired parameter and if the key P is pressed the display PV will show the paramete...

Страница 4: ...rogrammed on par SPLL and the one programmed on par SPHL Note in all the following examples the Set point is indicated as SP however the instrument will act according to the Set point selected as acti...

Страница 5: ...positive or negative offset that is simply added to the value read by the probe before visualisation which remains constant for all the measurements If instead it is desired that the offset set should...

Страница 6: ...ameters referring to the ON OFF control are contained in the group rEG This type of control can be obtained by programming par Cont On FS or On FA and works on the output programmed as 1 rEG depending...

Страница 7: ...at Power Ratio or relation between power of the element controlled by output 2rEG and power of the element controlled by output 1rEG 4 7 PID CONTROL FOR MOTORIZED ACTUATORS WITH TIME POSITIONING 1rEG...

Страница 8: ...grammed parameters To make the error ErAt disappear swap the instrument to the OFF control OFF and then turn it to automatic control rEG The Auto tuning cycle duration has been limited to 12 hours max...

Страница 9: ...ents etc where the set point has to be reached gradually in a predetermined time Once the instrument has reached the first Set Point SP1 it is possible to have automatic switching to the second Set Po...

Страница 10: ...band alarm OR MAXIMUM SET OF ALn ALARM THRESHOLD for low or high alarm HALn ALARM HYSTERESIS ALnd ALARM ACTIVATION DELAY in sec ALni ALARM BEHAVIOUR IN THE EVENT OF MEASUREMENT ERROR ALnt ALARM TYPE...

Страница 11: ...larm is active Enter group Hb and program which output the alarm signal has to address on parameter OHb The functioning mode of the alarm is instead defined on par HbF which can be set in the followin...

Страница 12: ...P1 2 By closing the contact connected to the digital input the set point SP2 is selected as being active while opening the contact selects the set point SP1 as active The function can only be activate...

Страница 13: ...l info please have a look at the KEY01 instruction manual 5 PROGRAMMABLE PARAMETERS Here following are described all the parameters available on the instrument Some of them could be not present or bec...

Страница 14: ...2 for high or low alarm AL2L 9999 9999 49 HAL2 Alarm AL2 hysteresis OFF 9999 1 50 AL2d Activation delay of alarm AL2 OFF 9999 sec OFF 51 AL2i Alarm AL2 activation in case of measuring error no yES no...

Страница 15: ...modes of the variable measured by the probe HCFG INPUT TYPE This permits selection of the input type thermocouples tc thermo resistances or thermistors rtd normalized signals in current I in voltage...

Страница 16: ...no_0 if one intends to use the beginning of the scale as being other than 0 4 mA or 2 V Ao1F ANALOGICAL OUTPUT OUT1 FUNCTION This determines the function of the OUT 1 output as control output 1 1 rEG...

Страница 17: ...LARM AL3 Similar to AL1d but referring to alarm AL3 AL3i ALARM AL3 ACTIVATION IN CASE OF MEASUREMENT ERROR Similar to AL1i but referring to alarm AL3 GROUP LbA PARAMETERS RELATIVE TO THE LOOP BREAK AL...

Страница 18: ...tched on the actuator must remain where it is no whether is must be taken to the maximum opening position oPEn or the maximum closing position cLoS Parameters relative to the ramps allowing the Set Po...

Страница 19: ...l stations must have the same transmission speed PACS ACCESS TO PROGRAMMING THROUGH SERIAL PORT Programming access If programmed as LoCL this means that the instrument can only be programmed from the...

Страница 20: ...ature 10 60 C 7 3 MECHANICAL DIMENSIONS PANEL CUT OUT AND MOUNTING mm 7 4 FUNCTIONAL FEATURES Control ON OFF single and double action PID PID for motorized actuators with time positioning Measurement...

Страница 21: ...99 9 999 9 19 99 99 99 1 999 9 999 0 10 V HCFG UoLt SEnS 0 10 1999 9999 199 9 999 9 19 99 99 99 1 999 9 999 2 10 V HCFG UoLt SEnS 2 10 1999 9999 199 9 999 9 19 99 99 99 1 999 9 999 7 6 INSTRUMENT ORDE...

Отзывы: