background image

 

TECNOLOGIC spa - TLK 43

 

- OPERATING INSTRUCTIONS - Vr.02 - ISTR 06267 - PAG. 6

 

= 2.rEG : Secondary control output 
= ALno : Alarm output normally open 
= ALnc : Alarm output normally closed 
= OFF : Output deactivated  
The coupling [outputs number – alarms number] can be effectuated 
in the group relative to the alarm (“

]

AL1”,”

]

AL2” or “

]

AL3”). 

- ANALOGICAL OUTPUTS 0/4..20 mA or 0/2..10 V (only OUT1,2):

 

The parameter 

“Aorn”

 will present within the group, with which it is 

possible to set the beginning of the scale used for the output. 
This parameter will therefore be set at: 
= 0 : if one intends to use the beginning of the scale as equal to 0 (0 
mA if the output is 0/4...20 mA, or 0 V if the output is  0/2...10 V) 
= no_0 : if one intends to use the beginning of the scale other than 0  
(4 mA if the output is 0/4...20 mA, or 2 V if the output is 0/2...10 V)

 

The parameter 

“AonF”

 will be present by which it is possible to 

configure the function of the analogical output as: 
= 1.rEG : Primary control output  
= 2.rEG : Secondary control output 
= r.inP : measurement retransmission output 
= r.Err : error retransmission output [SP-PV] 
= r.SP : Active Set Point retransmission output 
= r.SEr : output led by serial communication line of the instrument 
= OFF : deactivated output 
In the case that analogical output is configured as 1.rEG or 2.rEG 
the output signal  will be proportional to the control power calculated 
by the instrument starting from 0% (output signal corresponding to 
the set beginning of the scale) up to 100% (output signal 
corresponds to the maximum that can be supplied by the type of 
output available). 
The analogical control outputs can only be used for PID single 
action or dual action controls. 
If the set control mode was the ON/OFF type, the analogical output 
could only take on the control states 0 % or 100 %. 
In the case that the analogical output function should be configured 
for the retransmission of the signal, it is therefore necessary to 
programme another two parameters which set the minimum and 
maximum reference values. 
Therefore, in these cases, set the parameter 

"AonL"

 with the value 

that the instrument must provide the minimum value (0/4 mA or 0/2 
V) in output and the value to which the instrument must provide the 
maximum value (20 mA o 10 V) to the parameter

 "AonH"

 in output. 

 

4.3 – ON/OFF CONTROL (1rEG) 

All the parameters referring to the ON/OFF control are contained in 
the group 

]

rEG”. 

This type of control can be obtained by programming par.

"Cont"

 = 

On.FS or = On.FA  and works on the output programmed as 

1.rEG,

 

depending on the measure, on the active Set Point 

“SP”

, on the 

functioning mode 

"Func” 

and on the hysteresis 

"HSEt

". 

The instrument carries out an ON/OFF control with symmetric 
hysteresis if “Cont" = On.FS or with asymmetrical hysteresis if 
“Cont” = On.Fa. 
The control works in the following way : in the case of reverse 
action, or heating (“FunC”=HEAt), it deactivates the output, when 
the process value reaches [SP + HSEt] in case of symmetrical 
hysteresis, or [SP] in case of asymmetrical hysteresis and is then 
activated again when the process value goes below value [SP - 
HSEt]. Vice versa, in case of direct action or cooling ("Func”=CooL), 
it deactivates the output, when the process value reaches [SP - 
HSEt] in case of symmetrical hysteresis, or [SP] in case of 
asymmetrical hysteresis  and is activated again when the process 
value goes above value [SP + HSEt]. 

 

4.4 – NEUTRAL ZONE ON/OFF CONTROL (1rEG - 2rEG)

 

All the parameters referring to Neutral Zone ON/OFF control are 
contained in the group 

]

rEG”. 

This type of control can be obtained when 2 outputs are 
programmed respectively as 1rEG and 2rEG and the par. 

“Cont”

 = 

nr . 
The Neutral Zone control is used to control plants in which there is 
an element which causes a positive increase (ex. Heater, humidifier, 
etc.) and an element which causes a negative increase (ex. Cooler, 
de-humidifier, etc). 
The control functions works on the programmed outputs depending 
on the measurement, on the active Set Point 

“SP” 

and on the 

hysteresis 

"HSEt

". 

The control works in the following way : it deactivates the outputs 
when the process value reaches the Set Point and it activates the 
output 1rEG when the process value goes below value [SP - HSEt], 
or it activates the output 2rEG when the process value goes above 
[SP + HSEt].  
Consequently, the element causing a positive increase has to be 
connected to the output programmed as 1rEG while the element 
causing a negative increase has to be connected to the output 
programmed as 2rEG. 

 

 
4.5 – SINGLE ACTION PID CONTROL (1rEG) 

All the parameters referring to PID control are contained in the 
group 

]

rEG”. 

The Single Action PID control can be obtained by programming 
par.

"Cont"

 = Pid  and works on the output 1rEG depending on the 

active Set Point 

“SP”

, on the functioning mode 

"Func” 

and on the 

instrument’s PID algorithm with two degree of freedom. 

Содержание TLK 43

Страница 1: ...also instructions for its maintenance and use we therefore recommend that the utmost attention is paid to the following instructions Though this manual has been issued with the greatest care TECNOLOG...

Страница 2: ...ualized on 4 green displays while the outputs status is indicated by 4 leds The instrument provides for the storage of 4 Set Points and can have up to 4 outputs relay type or can drive solid state rel...

Страница 3: ...e visualised on the SV display by pushing the P key Again by using the UP and DOWN keys it is possible to select the desired parameter and if the key P is pressed the display PV will show the paramete...

Страница 4: ...rogrammed on par SPLL and the one programmed on par SPHL Note in all the following examples the Set point is indicated as SP however the instrument will act according to the Set point selected as acti...

Страница 5: ...positive or negative offset that is simply added to the value read by the probe before visualisation which remains constant for all the measurements If instead it is desired that the offset set should...

Страница 6: ...ameters referring to the ON OFF control are contained in the group rEG This type of control can be obtained by programming par Cont On FS or On FA and works on the output programmed as 1 rEG depending...

Страница 7: ...at Power Ratio or relation between power of the element controlled by output 2rEG and power of the element controlled by output 1rEG 4 7 PID CONTROL FOR MOTORIZED ACTUATORS WITH TIME POSITIONING 1rEG...

Страница 8: ...grammed parameters To make the error ErAt disappear swap the instrument to the OFF control OFF and then turn it to automatic control rEG The Auto tuning cycle duration has been limited to 12 hours max...

Страница 9: ...ents etc where the set point has to be reached gradually in a predetermined time Once the instrument has reached the first Set Point SP1 it is possible to have automatic switching to the second Set Po...

Страница 10: ...band alarm OR MAXIMUM SET OF ALn ALARM THRESHOLD for low or high alarm HALn ALARM HYSTERESIS ALnd ALARM ACTIVATION DELAY in sec ALni ALARM BEHAVIOUR IN THE EVENT OF MEASUREMENT ERROR ALnt ALARM TYPE...

Страница 11: ...larm is active Enter group Hb and program which output the alarm signal has to address on parameter OHb The functioning mode of the alarm is instead defined on par HbF which can be set in the followin...

Страница 12: ...P1 2 By closing the contact connected to the digital input the set point SP2 is selected as being active while opening the contact selects the set point SP1 as active The function can only be activate...

Страница 13: ...l info please have a look at the KEY01 instruction manual 5 PROGRAMMABLE PARAMETERS Here following are described all the parameters available on the instrument Some of them could be not present or bec...

Страница 14: ...2 for high or low alarm AL2L 9999 9999 49 HAL2 Alarm AL2 hysteresis OFF 9999 1 50 AL2d Activation delay of alarm AL2 OFF 9999 sec OFF 51 AL2i Alarm AL2 activation in case of measuring error no yES no...

Страница 15: ...modes of the variable measured by the probe HCFG INPUT TYPE This permits selection of the input type thermocouples tc thermo resistances or thermistors rtd normalized signals in current I in voltage...

Страница 16: ...no_0 if one intends to use the beginning of the scale as being other than 0 4 mA or 2 V Ao1F ANALOGICAL OUTPUT OUT1 FUNCTION This determines the function of the OUT 1 output as control output 1 1 rEG...

Страница 17: ...LARM AL3 Similar to AL1d but referring to alarm AL3 AL3i ALARM AL3 ACTIVATION IN CASE OF MEASUREMENT ERROR Similar to AL1i but referring to alarm AL3 GROUP LbA PARAMETERS RELATIVE TO THE LOOP BREAK AL...

Страница 18: ...tched on the actuator must remain where it is no whether is must be taken to the maximum opening position oPEn or the maximum closing position cLoS Parameters relative to the ramps allowing the Set Po...

Страница 19: ...l stations must have the same transmission speed PACS ACCESS TO PROGRAMMING THROUGH SERIAL PORT Programming access If programmed as LoCL this means that the instrument can only be programmed from the...

Страница 20: ...ature 10 60 C 7 3 MECHANICAL DIMENSIONS PANEL CUT OUT AND MOUNTING mm 7 4 FUNCTIONAL FEATURES Control ON OFF single and double action PID PID for motorized actuators with time positioning Measurement...

Страница 21: ...99 9 999 9 19 99 99 99 1 999 9 999 0 10 V HCFG UoLt SEnS 0 10 1999 9999 199 9 999 9 19 99 99 99 1 999 9 999 2 10 V HCFG UoLt SEnS 2 10 1999 9999 199 9 999 9 19 99 99 99 1 999 9 999 7 6 INSTRUMENT ORDE...

Отзывы: