background image

Cable fault locator

 

TDR-TA3.3T 

 

________________________________________________________________________ 

________________________________________________________________________ 

 

 

Table 1 

 

Probe 

pulse 

parameters  

10 

ns 

20 

ns 

50 

ns 

100 

ns 

200 

ns 

500 

ns 

1   

µs 

µs  5 µs 

10 

µs 

τ

и

µs 

≤ 

0,01 

≤ 

0,02 

≤ 

0,05 

0,1  ± 

0,01 

0,2  ± 

0,02 

0,5  ± 

0,05 

1,0  ± 

0,01 

2  ± 

0,2 

5  ± 

0,5 

10  ± 

1,0 

н

, ns,  

not more than 

10 

10 

10 

10 

10 

10 

10 

10 

10 

10 

U, 

В,  

not more than 

9,0 

9,0 

9,0 

9,0 

9,0 

9,0 

9,0 

9,0 

9,0 

9,0 

 

2.5 

Permissible  relative  error  parameters  of  the  shortening  coefficient within  the  range 

of 1 to 3 ± 0,84%. 

2.6 

Receiving path sensitivity  when the signal exceeds the noise  level twice  on all the 

subranges (windows) should be not less than 1 mW. 

2.7 

Agreed resistances range is from 30 to 400 Ohm. 

2.8 

Operation mode time set up is not more than 30 sec. 

2.9 

Continuous  running  TDR-TA3.3T  from  ACC1  battery  is  not  less  than  8  hours  and 

depends on the technical condition of battery. 

2.10  The TDR-TA3.3T overall dimensions are not more than: 

length 

– 230 mm 

width 

– 130 mm 

height 

– 50 mm 

2.11  The TDR-TA3.3T weight without removable batteries does not exceed 0,55 kg. 

2.12  Safety 

2.12.1 Average  operation  time  before  technical  resource  exhaustion  should  be  not  less 

than 6000 hours. 

2.12.2 Fixed service life should be not less than 5 years. 

 

 

Содержание TDR-TA3.3T

Страница 1: ...Cable fault locator TDR TA3 3T USER MANUAL...

Страница 2: ...operation test 15 6 2 Switching on of the device 15 6 3 Device connection to the researched line 15 6 4 Measuring the length of the researched line 16 6 5 Measuring the coefficient of line shortening...

Страница 3: ...___________________________________________ ________________________________________________________________________ 3 9 Device calibration 22 10 Transportation and storage 23 11 Marking 23 12 Accepta...

Страница 4: ...ics of the cable fault locator TDR TA3 3T Warranty terms do not cover the cases of misuse of the device or use of the device not in accordance with the rules listed in a given UM UM is intended for ac...

Страница 5: ...measuring the cable length measuring the distance to wave impedance irregularities or defects measuring the coefficient of a line shortening at its known length definition of nature of damage display...

Страница 6: ...s Measurement subranges windows 30 m 60 m 120 m 240 m 480 m 960 m 2400 m 4800 m 2 2 Distance measurement permissible basic reduced error limits time delay in the subranges windows 0 42 of the subrange...

Страница 7: ...of 1 to 3 0 84 2 6 Receiving path sensitivity when the signal exceeds the noise level twice on all the subranges windows should be not less than 1 mW 2 7 Agreed resistances range is from 30 to 400 Ohm...

Страница 8: ..._ ________________________________________________________________________ 8 3 Product and delivery set The TDR TA3 3T delivery set consists of domain reflectometer TDR TA3 3T 1 item connecting cable...

Страница 9: ...egularity type see Picture 2 The reflected pulses return to the device in some time from the moment of sending the probe pulse Knowing the electromagnetic wave propagation speed along the line and the...

Страница 10: ...point of a high resistance the second cursor that means a cable breakage The condition showed in the waveform is called after typical breakage The reflection with a signal polarity change is equivale...

Страница 11: ...ath The value of coefficient of shortening can be defined experimentally knowing the length of cable The error of the distance to irregularity is to be defined with discreteness of the indicator 320 d...

Страница 12: ...3 3T design 4 2 The control and displaying tools keyboard grips location and function 4 2 1 Power supply The power supply of the device is provided with four built in batteries AA 1 2 W 2700 A Hr The...

Страница 13: ...used for a quick switching on off of the device display s lightning see issue 6 12 4 2 8 The GAIN button is used for a quick transition to the Intensification selector for the gain factor adjustment s...

Страница 14: ...hortening window length of pulse The current parameters values are showed on the display 4 2 15 The and buttons in MEASUREMENT mode are used to move a measure cursor to the left or right 4 2 16 The di...

Страница 15: ...s acquainted with the given Manual are allowed to operate TDR TA3 3T 5 2 The TDR TA3 3T has no life threatening voltage 5 3 The staff operating on different tracks is obliged to follow the safety rule...

Страница 16: ...the device ATTENTION if the SECURITY button green indicator is not working the operation is prohibited because the researched line is under the electric voltage It is required to disconnect the device...

Страница 17: ...rms in the MEASUREMENT mode The reflected pulse of positive polarity marks the cable breakage It is necessary to set the cursor to the pulse forefront due to measure the distance The RI 303TM defines...

Страница 18: ...joints and breaks Picture 6 1 Sockets joints breaks Reflection from irregularity that distinguishes the cables joint is expressed in a positive polarity response The quality of a cable splice should...

Страница 19: ...ance to the sodden part of cable Picture 6 3 The sodden part of cable The irregularity connected with the appearance of moisture in the core of cable refers to extended irregularity The sodden part of...

Страница 20: ...he cable the pulse reflected from the end of cable might completely disappear The reflected pulse that by its shape indicates the availability of a parallel tap in the cable line is to be observed in...

Страница 21: ...rol in the TDR TA3 3T is carried out according to a pictograph in the top right corner of the display the symbol means the full charge of the batteries the symbol means that the batteries need to be c...

Страница 22: ...us can be assessed by the indicator in the upper left corner of the screen in MEASUREMENT mode 8 3 The device is equipped with a battery discharge control system see issue 6 11 8 4 Charge the battery...

Страница 23: ...ortation and storage 10 1 TDR TA3 3T packed in containers should be transported in closed vehicles of any type at any distance and stored under conditions established by a group of 4 to GOST 22261 Whe...

Страница 24: ...rviceable and found fit for use Release date ____ ______________ 20___ year LS The company representative __________________ _______________________ signature name Cable fault locator TDR TA3 3T seria...

Страница 25: ...acture 13 3 The warranty period of storage is set 6 months from the date of manufacture TDR TA3 3T 13 4 The manufacturer shall within the warranty period eliminate the identified defects or replace th...

Отзывы: