Peak Headings
Each peak cursor can be individually named by typing a name in the appropriate Peak Headings
box.
•
Click on the
PAR
button to view the
Peak Headings
box.
•
Type in a name for the peak.
The new name will then appear in the Results Table column heading and in the Report. This may
facilitate identification of peaks corresponding to different PCR products.
Threshold filter
The Threshold filter appears as a red horizontal sliding cursor across the dissociation peaks graph
at the default setting of 5%. This parameter limits the size of the peaks detected – peaks smaller
than the specified threshold % relative to the tallest peak are not recorded. For example, if a well’s
largest peak has a height of 1000 on the dissociation peak graph (-dF/dT vs. temp) and the
threshold is set to 5%, any peaks with a height of less than 50 are assumed to be noise and
ignored.
The cursor can be moved manually to
increase or decrease the threshold
value for peaks detected. The new
cursor position will be updated in the
PAR
box.
In a small number of cases, a Tm for a
peak may appear in the Results Table
even when the cursor is placed just
above peak. This is due to the
mathematical curve fitting algorithm
used to calculate the Tm. The peak
can be excluded by raising the cursor
slightly higher.
Содержание PrimeQ
Страница 1: ...Version 2 1 08 12 PrimeQ OPERATOR S MANUAL...
Страница 2: ...2...
Страница 5: ...5...
Страница 6: ...6...
Страница 12: ...12...
Страница 14: ...14...
Страница 40: ...40...
Страница 42: ...42...
Страница 52: ...52...
Страница 106: ...106...
Страница 108: ...108...
Страница 168: ...168...
Страница 170: ...170...
Страница 182: ...182...