142
4.7.2.4
Manual peak detect
To use manual peak detect, select this option on the Dissociation Wizard Peak Detector screen.
•
Number of peaks to find:
Up to a maximum of four.
•
Bin threshold:
The user can specify a temperature range (°C) in which a peak will be
considered valid.
•
Threshold filter:
The user sets a threshold value to limit the size of the peaks detected (see
below). The default is 5%.
Once the data has been collected, the specific number of cursors will be present on the
dissociation peak graph in the Results Editor.
To select further peaks:
•
Click on the
PAR
button next to the dissociation peak graph.
•
Enter the number of peaks required in the
Peak Detector
window of the Dissociation curve
parameters.
•
Click on each cursor using the mouse and drag each detection cursor to the desired position
on the graph.