7.
Remove fluid box adapter assembly (10) from the fluid manifold (4) by
using the nozzle adjustment tool to unscrew the locking screw (9).
Attention:
For the fluid manifold with heater, before using the nozzle
adjustment tool to unscrew the locking screw, first remove the cable
guide (11) so that it won’t interfere with the tool.
8.
Carefully remove the tappet seal from the tappet. Slightly squeeze both
sides of the seal (blue arrows) and slowly pull the seal from the tappet.
Caution:
Make sure that the seal does not become upended.
30
Jet Valve Dispensing System User Guide
TS9800
& TS980