19
4-2-4. F4 setup screens
x: Applicable
Setup
screen
number
Setup item name
Initial
setting
Setting range
Page
SET
Lock
CAL
Lock
Unit
Displayed text
F4-1
BCD Data Refresh Rate
100
1, 2, 5, 10, 20, 50, 100
56
x
# of
times
/sec
F4-2
BCD Output Check
C bcd
56
F4-3
RS-232C Interface Setting
13100
1 3 1 0 0
2
2
2
2
2
a b c d e
a Transmission mode
0: Command transmission mode
1: Continuous transmission mode
b Baud rate
0: 1200bps
1: 2400bps
2: 4800bps 3: 9600bps
4: 19200bps
c Data bits
0: 7bits
1: 8bits
d Parity bits
0: None
1: Odd
2: Even
e Stop bits
0: 1bit
1: 2bits
58
x
F4-4
DA Zero
000.00
-19999 to 99999
67
x
F4-5
DA Full Scale
100.00
-19999 to 99999
67
x
F4-6
Save TEDS Data
2.000
70
x
F4-7
Restore TEDS Data
00000
71
x
F4-8
Decision-Output Check
C Out
Turn on the decision outputs for HH, HI,
OK, LO, and LL, one by one
41
F4-9
Control-Input Check
C IN
Show the ON/OFF state of CLEAR, HOLD,
JUDGE, and DZ on the LED display
41
F4-10
Beep
ON
OFF, ON
42
x
F4-11
Select Data Output
0.dSP
0.dISP: Reference the display; hold value
is sent in hold mode
1.InP: Reference the input; input value is
sent in hold mode
42
x
4. Setup Mode