Z2000/C2020 Tank Gauge Setup Guide – PD02/0001 Rev.01 Feb/2017

 

Z2000 / C2020 

Tank Gauge Set-Up

 

 

To Calibrate the Gauge, Open the front cover 4 screws, 

Connect the Tank  

Probe, and Power the unit

.         

Toggle the calibration switch, top right marked CAL (LK2) 

       
 

The first screen to show is a “Cube Tank”     Middle Button Scrolls through Tank Type Opti

ons 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Hidden Button Scrolls through settings. 

 

 

 

Hidden Button 

Choose the Feature you require, then set the 
values by using the hidden button to select the 
value to change, then use the other two buttons 
to Decrease (LEFT) or Increase (RIGHT)the value. 

You Must Scroll to YES and use the 

Hidden

 

button to 

Confirm 

YES 

for all Screens

 

 

Now Use the 

Right Button

 

to Scroll across to 

Alarm

 Trigger points and 

SG

 setting 

 

You must scroll to YES and 
Confirm YES with Left 
hand Hidden Button to 
save your values. 

Left (hidden Button) selects Option          
Right two buttons change value as in 5 
above. 

 