25
6-1-5. Display Refresh Rate
You can select the display refresh rate per second for a reading.
1. Enter the setup screen number F3-4.
Press 3 times
Press 3 times
2. All the digits of the displayed value blinks.
To select 4, 6, 10, or 20, press
5
or
b
.
or
3. To save the selection, press ENTER, and then to exit
the setup mode, press ESC.
To save, press:
NOTE
The initial setting is 4.
6-2. Equivalent input calibration
You can determine the calibration values by setting the rated
output and rated capacity shown on the test report.
Using this calibration method, you can easily perform a
calibration without applying an actual load.
Set rated output
Set rated capacity
Zero-point calibration
1. Press CAL, and then press ENTER.
The rated output (mV/V) appears on the display with the
least significant digit blinking.
2. To set the rated output;
change the blinking digit of the value by pressing
5
or
b
.
To move to the previous or next digit, press
g
or
t
.
To change the digit, press:
To move to the previous
or next digit, press:
3. To save the rated output you entered, press ENTER.
The rated capacity appears with the least significant digit
blinking.
6. Calibration