![Teac TD-240A Скачать руководство пользователя страница 98](http://html1.mh-extra.com/html/teac/td-240a/td-240a_operation-manual_827107098.webp)
16.CONFORMITY TO EC DIRECTIVES
89
16. CONFORMITY TO EC DIRECTIVES
The TD-240A digital indicator conforms to EC directives (based on the EC council of
ministers), carrying a CE mark.
- Low voltage directive:
EN61010-1
- EMC directive:
EN61326-1
When installing, attention should be given to the following.
1. Since the TD-240A is defined as an open type (built-in equipment), be sure to install
the TD-240A and fix to a panel or the like for use.
2. Use the attached power cable.
3. Use shielded cables for others (load cell, external I/O, option).
Attachment of a ferrite core
It is necessary to twist a power supply cable and sensor cables, such as a load cell,
around an attached ferrite core.
(
Common to AC spec. and DC spec.
)
EN55011
,
EN61000-4-2
,
EN61000-4-3
,
EN61000-4-4
EN61000-4-5
,
EN61000-4-6
,
EN61000-4-8
EN61000-4-11
,
EN61000-3-2
,
EN61000-3-3
Load cell
A cable is twisted around
a ferrite core.
(
One roll
)
TD-240A
Ferrite core
power supply cable