4.CALIBRATION
29
・
Actual Load Calibration
If a calibration error display appears, take a proper action according to the error, then
perform calibration again.
…
Span set value is "00000".
Set a correct span value.
…
Output of the strain gauge sensor does not reach the span adjustment
range.
Confirm whether an actual load is put on the strain gauge sensor.
Calibration may not be performed without load
…
Output of the strain gauge sensor is on the minus side.
Check to see if the +SIG and -SIG wiring if the sensor is reversed.
On completion of calibration, turn on the calibration LOCK.
1) Apply an actual load to the sensor and set the actual load value.
Use
and
keys to set the actual load value
,
then
key to validate the setting.
To move the decimal point, press
key at a blinking digit.
If the indicated value following the display CALSP is
Actual Load, the Actual Load calibration is terminated.
ENT
SEL
UP
DOWN
SEL
ENT
FNC
Actual Load Calibration
Indicated Value of
(
00000 to 19999
)
Actual Load Value