5.SETTING OF FUNCTIONS
40
5-7. Analog Filter
This is a lowpass filter filtering the strain gage sensors input signal and cutout the noise
element.
Lowpass filter cutout freguency is selectable in the 4/10/100/3k Hz.
1) Select setting mode 2.
2) Select analog filter.
Use
and
keys to set the analog filter,
then use
key to validate the setting.
To return to the indicated value display, press
key.
FNC
SEL
DOWN
Press twice.
ENT
UP
DOWN
ENT
ESC
Setting of Analog Filter
3
:
3kHz
2
:
100Hz
1
:
10Hz
0
:
4Hz
Analog Filter