![Teac TD-240A Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/teac/td-240a/td-240a_operation-manual_827107014.webp)
1.FUNCTIONAL DESCRIPTIONS
5
1-2-5. Signal Input/Output Terminal Board
This terminal board is used for input/output of control signals, SI/F data output, and
input of strain gauge sensor signals.
・
Terminal board Assignment
1
・
2
:
Two-wire serial interface
(
SI/F
)
for connecting printers and external display
from TEAC.
This interface has no polarity and can connect up to three external devices.
Use parallel two-core cables or captire cables.
3
~
6
:
Output terminals of the high/low limit relays.
3
・・・
High limit relay
(
N/O
)
4
・・・
High limit COM
5
・・・
Low limit relay
(
N/O
)
6
・・・
Low limit COM
Rating is 250V AC and 0.5A.
㧿㧱㧾㧚㧺
10
11
12
13
14
15
SER.N
1
11
12
13
14
15
2
3
4
5
6
7
8
9
10
SI/F
N/O
COM
High limit relay
Low limit relay
Hold input
COM
DZ input
Analog voltage output
Analog COM
+
EXC
-
SIG
-
EXC
+
SIG
SHIELD
N/O
COM
Strain gauge
Connect the shield
cable of the sensor.
sensor input
Содержание TD-240A
Страница 1: ...TD 240A OPERATION MANUAL TEAC INSTRUMENTS CORPORATION Rev 1 35 2005 JUL ...
Страница 89: ...13 SELF CHECK FUNCTION AND INITIALIZATION 80 Self check Visual Check Sequence ...
Страница 91: ...13 SELF CHECK FUNCTION AND INITIALIZATION 82 Initialization Sequence ...