![Teac TD-240A Скачать руководство пользователя страница 68](http://html1.mh-extra.com/html/teac/td-240a/td-240a_operation-manual_827107068.webp)
8.BCD DATA OUTPUT
(
TD-2403
)
59
8. BCD DATA OUTPUT
(
TD-2403
)
The BCD Data Output Interface is for transferring indication values in BCD (Binary
coded Decimal) from to PC's PLC's or sequences for controlling, processing and
recording data.
The internal and external circuits are opto-isolated.
8-1. Connector Pin Assignment
Amphenol Connector
(
36-Pin
)
Compatible connector is DDK57-30360 or equivalent.
No.
Signal
No.
Signal
1
*
COM
19
*
COM
2
Out
1
20
3
Out
2
21
4
Out
4
22
Out
Near Zero
5
Out
8
23
Out
Minus
(
polarity
)
6
Out
10
24
Out
OVER
7
Out
20
25
Out
P.C
(
stable
)
8
Out
40
26
Out
STROBE
9
Out
80
27
In
BCD Data hold
10
Out
100
28
In
Logic Switching
11
Out
200
29
12
Out
400
30
13
Out
800
31
14
Out
1000
32
15
Out
2000
33
16
Out
4000
34
17
Out
8000
35
18
Out
10000
36
Содержание TD-240A