5.SETTING OF FUNCTIONS
44
・
The Zero Tracking function automatically resets the zero point to zero at a specified
time interval when the zero point move amount is below the specified range.
・
The time
(
tracking delay
)
is set in units of 0.1 seconds, from 0.1 to 9.9 seconds. The
range
(
tracking band
)
is set in units of quarters of the indicated value.
(
The indicated
value 02 corresponds to 0.5 scales and 12, 3 scales.
)
If the time is set to 0.0 sec. and
range to 00, the Zero Tracking function does not work
.
Boundary of zero track
BAND=COUNT
×
the minimum digital scale
×
2
Indicated Value
DELAY
+COUNT
-COUNT
BAND
+
0
-
DELAY
From the point when it returned within the range,
counting will be resumed.
The Zero Tracking works from where the indicated value is zero. It does not work when
the indicated value exceeds the tracking band. In this case,specify the zero point using
the Digital Zero or Zero Calibration.
Request