Terminal Block Connections Optional Equipment Wiring
Illustration 1: Optional Equipment Connections
Terminal Block Connections Optional Equipment Wiring. An oscilloscope, function generator
and remote DC ON/OFF switch are shown installed to the Dynaload. The terminal block may
also be used to connect various optional equipment to the Dynaload. In the example shown in
figure 10.
Note: All XBL’s have a, factory installed, jumper across E- and S- to
enable “Load On”. A normally closed emergency cut off switch can be
connected at this point.
XBL Series 404078 Rev P2.odt Page 19 Of 162
Caution: External Instruments connected to the (TB) terminal
block should be isolated
not
earth grounded. All input or output
TB connections should be referenced to S--. Emergency Power
Off (EPO) control is normally closed.
( )
( )
Remove (E+) and (E-) jumpers
for remote voltage sensing.
Connect to source
C
S
R
E
M
S
Y
N
E
N
S
–
S
–
E
–
S
+
E
+
402825-02-11.CDR
Oscilloscope
Function Generator
Remote DC
ON/OFF
Switch