
N08A29;15/33
- 15 -
Table 6-2 Start-up condition for each setting Group
Start-up condition (at time of power-on)
The groups of Operation
Mode Setting
Operation state of camera at
start-up
Position of mode switch
UP/DOWN switch
Arbitrary position
from 0 to F
Neutral position (no switch
manipulation)
Group 1
By normal automatic loading
(
→
Table 6-8)
Table 6-3 Set contents for Operation Mode Setting [Group 1]
Position of
mode switch
Item to be changed
UP/DOWN switch
UP position
DOWN position
0
Change of gain
Increase in gain
Decrease in gain
1
Menu display
(MENU)
<* C>
Group2
MENU OFF
2
Selection of gain preset 1 or 2
Preset 2 (
±
0db)
Preset 1 (-3db)
3
Selection of gain preset 3 or 4
Preset 4 (+12db)
Preset 3 (+6db)
4
(Not used)
5
Shutter – continuous/asynchronous (S.FORM )
Continuous
(NORMAL)
Asynchronous
(ASYNC)
6
Shutter – high speed/low speed
(S.TIME)
High speed (HIGH)
Low speed (LOW)
7
Change of shutter speed (S.TIME)
Shorter
Longer
8
Digital offset
(OFFSET)
Increase in offset
Decrease in offset
9
Switch between normal scan/partial scan (SCAN)
All pixels (NORMAL)
Partial (PARTIAL)
A to F
Program pages from A to F
Save
Load
(Note ) Decibel value in “Selection of gain preset” is a relative value based on Preset 1.
(Note) The shutter speed can be changed only when the shutter switch is currently set at any other positions than “0”.
When the shutter switch is positioned at “0”, the shutter speed cannot be changed even if the mode switch is set to
the position “7” and the UP/DOWN switch is manipulated. (The shutter is always OFF where shutter switch = “0”.)
Table 6-4 Set contents for Operation Mode Setting [Group 2]
Position of
mode switch
Item to be changed
UP/DOWN switch
UP position
DOWN position
1
Menu display
(MENU)
<* C>
Group C
MENU OFF
2
Gain correction (GAIN B)
+Gain correction
-Gain correction
3
OFFSET correction (OFFSET B)
+Offset correction
-Offset correction
4
Enabled/disabled for pulse width control (PWC) (!)
ENABLED
DISABLED
A to F
Program pages from A to F
Save
Load
(!) When the setting of “ Enabled/disabled for pulse width control” is not set to be “ENABLED”, the pulse width control is
not effective even if the shutter switch is set to the position “9” for the asynchronous shutter operation. In this case, the
preset shutter in asynchronous shutter mode is effective.
Table 6-5 Set contents for Operation Mode Setting [Group C]
Position of
mode switch
Item to be changed
UP/DOWN switch
UP position
DOWN position
1
Menu display
(MENU) <* C>
Group D
MENU OFF
2
Operation confirmation buzzer
(BZ)
<* C>
ON
OFF
3
Serial communication baud rate (BAUDP) <* C>
19200 bps
9600 bps
4 to 5
(Not used)
(-)
-
-
6
Polarity setting for Vinit2
(Vinit2) <* C>
Positive logic
(INVERTED)
Negative logic
(NORMAL)
7
Continuous shutter strobe signal (STRB-C) <* C>
ON
OFF
Table 6-6 Set contents for Operation Mode Setting [Group 4]
Position of
mode switch
Item to be changed
UP/DOWN switch
UP position
DOWN position
1
Menu display
(MENU)
<* C>
Factory default
MENU OFF
2
Test pattern
(PATTERN)
<* C>
ON
OFF
3
Output bit
(BIT)
<* C>
10bit
8bit
4 to 6
<Not used>
(-)
-
-
7
Vsub voltage
(VSUB)
<* C>
Increase in voltage
Decrease in voltage
•
(Note)
Vsub voltage is appropriately adjusted before shipment, and therefore the setting value should
not be changed
unless there is some particular reason
.
If the value is changed by mistake, it can be reset to the factory default by executing “(6-9) Read out of factory
default”.
[Explanation] Vsub voltage
Vsub voltage is the bias voltage (substrate voltage) that serves to control the blooming effect (resulting in blur or running
image at a saturating area) that arises from excessive light getting into CCD.
If a high Vsub voltage is set, the blooming effect can be reduced, although an excessively high voltage leads to a
narrower operation range of CCD because it is associated with a decrease in the saturating voltage of the CCD output.
It is appropriately set before shipment because the optimum Vsub voltage varies by CCD.