TP350 User’s Manual
3. Ramp Rate Test
①
From the ‘home’ screen the ‘PCR File Manager’ icon will bring to the
experiment page. You can load an existing experiment from the internal drive.
②
Click the ‘Dice’ icon from some accounts
③
Select the Ramp Rate Test.t3p file to preset program in ‘Dice’ and click the
checkmark
④
Touch a ‘Edit File’ icon, show the content of program.
Program details
:
25
℃
for 30sec, 95
℃
for 30sec, 40
℃
for 30sec (1 cycle)
⑤
Activating ’Run File’ button
under the icons will show a profile review page
which allows the operator to check over the profile (displayed in text format)
before confirming commencement and then click ‘Start Run’.
v2.01
33