
TP350 User’s Manual
●Chapter 5: Diagnosis Functions
The following are provided in three test files to program of ‘Dice’ account folder.
Cycle Test
:
Test cycling time and stability of control operations
Up Down Test
:
Measures time required from heat up to 98
℃
to cool down
to 20
℃
Ramp Rate Test
:
Tests heating and cooling rate
【
Caution
】
Confirm a standby temperature is set at room temperature prior to commencing test.
The Diagnosis function performs the following three tests
1. Cycle Test
①
From the ‘home’ screen the ‘PCR File Manager’ icon will bring to the
experiment page. You can load an existing experiment from the internal drive.
②
Click the ‘Dice’ icon from some accounts
③
Select the Cycle Test.
t
3
p
file to preset program in ‘Dice’ and appear the
specification of this program at the right side of PCR File Manager.
Touch a ‘Edit File’ icon, show the content of program.
Program details
:
94
℃
for 30sec, 55
℃
for 30sec, 72
℃
for 30sec (1 cycle)
v2.01
29