Pag. 28 USER MANUAL T-ONE Medi/Medi-Pro cod. doc. MNPG06-03
Settable values chart:
PARAMETER DESCRIPTION
MIN
MAX
Frequency
1 Hertz
120 Hertz
Impulse width
20 microsec.
450 microsec.
Total time
1 minute
60 minutes
M03 • EMS 1 second slope
.
EMS 1 sec. slope:
to set
EMS program with 1 second impulse slope.
•
Use keys
to move through the menu and select the stimulation parameter, use keys
to modify the values
•
Set all stimulation parameters, position the cursor over OK to store the program
•
To start the stored program, press key
.
•
A three sound signals will advise of the end of the program.
Settable values chart:
PARAMETER DESCRIPTION
MIN
MAX
Contraction frequency
20 Hertz
120 Hertz
Contraction time
2 seconds
10 seconds
Recovery frequency
1 Hertz
12 Hertz
Recovery time
2 seconds
20 seconds
Impulse width
60 microseconds 450 microseconds
Total time
1 minute
60 minutes
IMPORTANT:
using 1 second slope the contraction will be faster, while with 2/3
seconds slope we obtain more progressive contraction. Particularly for memory 5, with 3
seconds slope, we obtain a “triangular” impulse where the impulse parameters are
continuously varied. This program is particularly suitable for partially denervated muscles.