Pressing the « Calibration » button opens the programming window.
Offset
Gain
Erasing
Calibration principle is identical for pH measurement or chlorine measurement.
-
To calibrate the pH in 1 point on a photometer reading, use the
Offset
tab
.
-
If the calibration is done at pH7 we will use the
Offset
tab
.
-
If the calibration is done at pH4 or 9 we will use the
Gain
tab
.
-
To calibrate the chlorine in 1 point on a reading photometer we will use the
Gain
tab
.
-
Offset tab of chlorine is very rarely used unless requested by an experienced technician.
-
To return to the "factory" version, use the
Clear
tab
.
➢
Offset or pH7 calibration
Select the Offset tab.
Press the key to open the input keyboard of the new near pH 7.
Press calibrate to validate the entry.
-
In case of correct calibration "Calibration OK" is displayed in green
-
In case of calibration error "Error Calibration" is displayed in red
➢
Gain or pH4 calibration
Select the Gain tab.
Press the key to open the input keyboard of the new near pH 4.