Troubleshooting
7.
Troubleshooting
This chapter provides some hints to make troubleshooting easier.
For any detailed information on how to handle/clean parts please
see chapter
For any detailed information on how to program the instrument
please see chapter
Program List and Explanations, p. 50
.
If you need help please contact your local distributor. Note serial
number of instrument and all diagnostic values before.
Conditions for
pH calculation
only 1 alkalization agent (acid-base pair) in the sample (no
mixture)
the contamination is mostly NaCl
phosphate concentration is < 0.5 ppm
if pH value is < 8, the concentration of contaminant must be
small compared to the concentration of the alkalization agent
pH value is > 7.5, and < 11.5
Problem
Possible Reason
Cond. value
<0.055
S/cm
Air bubble at sensor tip or sensor in air.
No pH/alkaliza-
tion agent value
available in dis-
play, relay, signal
output
Switch on calculations in <Installation> /
<Sensor> / <Miscellaneous> /
<Calculations>.
Afterwards program screen 1 and 2 in
<Operation> / <Display> / <Screen 1>,
<Screen 2>.