Installation
3.4.
Electrical Connections
WARNING
Electrical hazard.
Always turn off AC power before manipulating electric
parts.
Grounding requirements: Only operate the instrument
from an power outlet which has a ground connection.
Make sure the power specification of the instrument corre-
sponds to the power on site.
Cable
thicknesses
In order to comply with IP66, use the following cable thicknesses
NOTICE:
Protect unused cable glands
Wire
For Power and Relays: Use max. 1.5 mm
2
/ AWG 14
stranded wire with end sleeves.
For Signal Outputs and Input: Use 0.25 mm
2
/ AWG 23
stranded wire with end sleeves.
A
B
C
PG 11 cable gland: cable Ø
outer
5–10 mm
PG 7 cable gland: cable Ø
outer
3–6.5 mm
PG 9 cable gland: cable Ø
outer
4–8 mm
A
B
C