10. Default Values
Operation:
Sensors:
Filter Time Const.: .................................................................20 Sec
Hold after Cal.:......................................................................... 0 sec
Relay Contacts
Alarm Relay .................................................. same as in Installation
Relay 1/2...................................................... same as in Installation
Input.............................................................. same as in Installation
Logger:
Logger Interval:.............................................................. 30 Minutes
Clear Logger:................................................................................ no
Display:
Screen 1 and 2; Row 1: ................................................. Cond 1(sc)
Screen 1 and 2; Row 2: .................................................. Cond 2(cc)
Screen 1 and 2; Row 3: ........................................................... None
Installation:
Sensors
Miscellaneous; Calculations: ........................................................ no
Miscellaneous; Meas. unit ..................................................... µS/cm
Sensor Parameters; Sensor 1 and 2;Cell Constant......0.0415 cm
-1
Sensor Parameters; Sensor 1 and 2; Temp. corr. ................ 0.00 °C
Sensor Parameters; Sensor 1 and 2; Cable length ................. 0.0 m
Sensor Parameters; Sensor 1; Temp. comp.; Comp: ........Ammonia
Sensor Parameters; Sensor 2; Temp. comp.; Comp: ...Strong Acids
Signal Output 1
Parameter:...................................................................... Cond 1(sc)
Current loop:..................................................................... 4 –20 mA
Function:.................................................................................. linear
Scaling: Range low:............................................................0.000 µS
Scaling: Range high: ......................................................1000.00 µS
Signal Output 2
Parameter:...................................................................... Cond 2(cc)
Current loop:..................................................................... 4 –20 mA
Function:.................................................................................. linear
Scaling: Range low:............................................................0.000 µS
Scaling: Range high: ......................................................1000.00 µS
Alarm Relay:
Conductivity; Cond. 1 (sc) and Cond. 2 (cc):
Alarm high: .....................................................................3000.00 µS
Alarm low:...........................................................................0.000 µS
Hysteresis:............................................................................10.0 µS
Delay: ......................................................................................5 Sec
Sample Temp: (Temp. 1 and Temp. 2)
Alarm High:............................................................................ 160 °C
Alarm Low:................................................................................. 0 °C