Program List and Explanations
A-96.250.871 / 120417
55
5 Installation
5.1 Sensors
5.1.1
Miscellaneous
5.1.1.1
Calculations:
Select “yes” if pH and ammonia concentrations
should be calculated. pH and ammonia are now available on
screen 1 or 2, on the signal outputs and as alarm or limit values.
5.1.1.2
Meas. unit:
Choose the measuring unit as µS/cm or µS/m
5.1.2
Sensor parameters
5.1.2.1
Sensor 1
5.1.2.1.1
Cell Constant
: Enter the cell constant printed on the sensor label.
Range: 0.0300 cm
-1
to 0.0600 cm
-1
5.1.2.1.2
Temp. Corr
: Enter the temperature correction printed on the sensor
label.
Range: -1 °C to 1 °C
5.1.2.1.3
Cable length
: Enter the cable length. If the flow cell is installed on
the monitor, set the cable length to 0.0 m.
Range: 0.0 m to 30.0 m
5.1.2.1.5
Temp. comp
5.1.2.1.5.1
Comp
.: Available compensation models:
Strong acids (Never select strong acids for sensor 1!)
Strong bases
Ammonia
Morpholine
Ethanolamines
Neutral salts
High purity water
Coefficient
none
5.1.2.2
Sensor 2
5.1.2.2.1
Cell Constant
: Enter the cell constant printed on the sensor label.
Range: 0.0300 cm
-1
to 0.0600 cm
-1
5.1.2.2.2
Temp. Corr
: Enter the temperature correction printed on the sensor
label.
Range: -1 °C to 1 °C
5.1.2.2.3
Cable length
: Enter the cable length. If the flow cell is installed on
the monitor, set the cable length to 0.0 m.
Range: 0.0 m to 30.0 m