![Swann AMI CACE Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/swann/ami-cace/ami-cace_manual_1411469014.webp)
12
A-96.250.871 / 120417
AMI CACE
Product Description
2.2.
Instrument Specification
Power Supply
Voltage:
100–240 VAC (± 10%)
50/60 Hz (± 5%)
or 24 VDC (± 10%)
Power consumption:
max. 30 VA
Sample
requirements
Flow rate:
3–4 l/h
Temperature:
up to 50 °C
Inlet pressure:
up to 0.5 bar
Outlet pressure:
pressure free
Use of a SWAN Backpressure Regulator is highly recommended.
Particle filtration recommended in case of high iron concentration.
NOTICE:
No oil, no grease, no sand. Use of film forming
products may reduce lifetime of EDI module.
On-site
requirements
The analyzer site must permit connections to:
Sample inlet:
Swagelok 1/4” adapter for stainless
steel tube
Sample outlet:
G 3/8” adapter for flexible tube
diam. 20 x 15 mm
Measuring
range
Measuring range
Resolution
0.055 to 0.999
S/cm
0.001
S/cm
1.00 to 9.99
S/cm
0.01
S/cm
10.0 to 99.9
S/cm
0.1
S/cm
100 to 1000
S/cm
1
S/cm
Automatic range switching.
EDI capacity
SC
max
= 40
S/cm as NH
4
OH
SC
max
= 350
S/cm as NaOH
Accuracy
±1% of measuring value or ±1 digit (whichever is greater)
Electronics
housing
Aluminum with a protection degree of IP 66 / NEMA 4X
Ambient temperature:
-10 to +50 °C
Humidity:
10–90% rel., non condensing
Display:
backlit LCD, 75 x 45 mm
Dimensions
Panel: Dimensions:
280 x 850x 200 mm
Screws:
8 mm diameter
Weight:
14 kg