![Swan Analytical Instruments A-85.119.010 Скачать руководство пользователя страница 59](http://html1.mh-extra.com/html/swan-analytical-instruments/a-85-119-010/a-85-119-010_manual_1411374059.webp)
6 - Maintenance - Repair
Calibration
Standard pH
To prevent the instrument from starting compensatory
regulation upon removing the pH sensor from the overflow cell:
Š
on the
Calibration
menu, choose
Standard pH
Š
press Enter
Š
only now remove the pH sensor from the overflow cell
and carry out calibration.
Calibration
Process pH
3.1.2
Standard pH
3.1.2.5
Standard pH
Zero Trides
Process Trides
Enter
Rinse and dry Electrode
and put it in
Standard 1
<Enter> to continue
Note:
Standard pH
3.1.2.1
If the pH value is outside the
expected range, check the
calibration solution. Besides the
loss of quality due to frequent
usage, service life of the
solution is limited:
Standard 1
7.00 pH
Current Value
x.xx pH
Offset
x.xx mV
Progress
<Enter> to save
- 12 months if kept at 5 °C
Standard pH
3.1.2.5
- 6 months if kept at 25 °C
For best results, make sure to
use fresh calibration solution
pH7/pH9. See
10.
4 “Consumables”,
123
for
Rinse and dry Electrode
and put it in
Standard 2
<Enter> to continue
Standard pH
3.1.2.1
ordering information.
Standard 2
9.00 pH
Current Value
x.xx pH
Slope
x.xx mV
Progress
<Enter> to save
Standard pH
3.1.2.5
Rinse and dry Electrode
and put it into
the Flow cell
<Enter> to continue
AMI Trides Operator’s Manual, A-96.250.101
51
Содержание A-85.119.010
Страница 10: ...1 AMI Trides keys 2 AMI Trides Operator s Manual A 96 250 101...
Страница 15: ...2 Safety Instructions General 2 2 1 AMI Trides Operator s Manual A 96 250 101 7...
Страница 16: ...2 Safety Instructions Warning Notices 2 2 corrosive 8 AMI Trides Operator s Manual A 96 250 101...
Страница 18: ...2 2 3 5 3 5 5 Dimethyldantoin 2 3 6 relay 1 relay 2 2 3 7 ESD 10 AMI Trides Operator s Manual A 96 250 101...
Страница 20: ...3 3 1 2 AMI Trides pH 3 2 3 12 AMI Trides Operator s Manual A 96 250 101...
Страница 22: ...3 Structure and Function Hardware 3 2 AMI Trides LCD PH 14 AMI Trides Operator s Manual A 96 250 101...
Страница 25: ...3 Structure and Function Hardware 6 mm inside dia 14 mm 3 PH pH AMI Trides Operator s Manual A 96 250 101 17...
Страница 32: ...3 Structure and Function Software 24 AMI Trides Operator s Manual A 96 250 101...
Страница 40: ...5 Diagnostics Menu Diagnostics 2 32 AMI Trides Operator s Manual A 96 250 101...
Страница 64: ...6 Maintenance Repair Calibration 56 AMI Trides Operator s Manual A 96 250 101...
Страница 76: ...7 Operation Logger 4 4 optional 68 AMI Trides Operator s Manual A 96 250 101...
Страница 123: ...8 Installation Instrument Setup AMI Trides Operator s Manual A 96 250 101 115...
Страница 128: ...9 Shut Down Transport Disposal Disposal 120 AMI Trides Operator s Manual A 96 250 101...