4 - Error Messages
Menu Messages 1
4
Error Messages
4.1 Menu Messages 1
The menu Messages 1 reveals pending errors as well as
message text, time and state of errors that have occurred
before.
OFF
R1
R2
14:10:45
0.42 ppm
7.04 pH
Error
Nonfatal error. The system continues
to operate normally, except for the
defective part.
157 rpm
Enter
25.
Main
Menu
Messages
Diagnostics
Maintenance
Operat
En
ion
ter
4°C
Messages
1
1.1
Fatal error
(blinking symbol)
Control of dosing devices is
interrupted.
The indicated measured values are
Installation
Pending Errors
Message List
incorrect.
Pending Errors
1.1.5
Errors remain in the
Enter
Error Code
E010
Sample Flow low
<Enter> to Acknowledge
Pending Errors window until
corrected, even if
acknowledged.
Acknowledging the error
only deactivates the Alarm
Relay, i.e. it switches off an
alarm device, if available.
Messages
Pending Errors
Message List
1.2
If an error is corrected, by
intervention or by the
system itself, it is listed in
the
Message List 1.2
, the
Enter
Message List
1.2.1
Number
27
Date, Time 15.07.03 08:52:27
E010
Sample Flow low
State
Came in
symbol
disappears.
The last 64 messages are
stored for viewing. They are
deleted if
Set Default Values
is carried out.
AMI Trides Operator’s Manual, A-96.250.101
25
Содержание A-85.119.010
Страница 10: ...1 AMI Trides keys 2 AMI Trides Operator s Manual A 96 250 101...
Страница 15: ...2 Safety Instructions General 2 2 1 AMI Trides Operator s Manual A 96 250 101 7...
Страница 16: ...2 Safety Instructions Warning Notices 2 2 corrosive 8 AMI Trides Operator s Manual A 96 250 101...
Страница 18: ...2 2 3 5 3 5 5 Dimethyldantoin 2 3 6 relay 1 relay 2 2 3 7 ESD 10 AMI Trides Operator s Manual A 96 250 101...
Страница 20: ...3 3 1 2 AMI Trides pH 3 2 3 12 AMI Trides Operator s Manual A 96 250 101...
Страница 22: ...3 Structure and Function Hardware 3 2 AMI Trides LCD PH 14 AMI Trides Operator s Manual A 96 250 101...
Страница 25: ...3 Structure and Function Hardware 6 mm inside dia 14 mm 3 PH pH AMI Trides Operator s Manual A 96 250 101 17...
Страница 32: ...3 Structure and Function Software 24 AMI Trides Operator s Manual A 96 250 101...
Страница 40: ...5 Diagnostics Menu Diagnostics 2 32 AMI Trides Operator s Manual A 96 250 101...
Страница 64: ...6 Maintenance Repair Calibration 56 AMI Trides Operator s Manual A 96 250 101...
Страница 76: ...7 Operation Logger 4 4 optional 68 AMI Trides Operator s Manual A 96 250 101...
Страница 123: ...8 Installation Instrument Setup AMI Trides Operator s Manual A 96 250 101 115...
Страница 128: ...9 Shut Down Transport Disposal Disposal 120 AMI Trides Operator s Manual A 96 250 101...