8 - Installation
Instrument Setup
Function
Control upw., dnw.
Parameter
DIS.
Actuator
Time proportional
Relay 1
5.3.2.1
recommended:
Relay 1
DIS
Function
Control upw.
Parameter
DIS.
Relay 2
pH
Settings
Settings
5.3.2.32.1
Actuator
Actuator Time proportional
Cycle Time
60 Sec
Response Time
10 Sec
Time proportional
Frequency
Motor Valve
Control Parameters
Setpoint
default for
Relay 2
is
10.00 ppm
Control Parameters
5.3.2.32.4.10
Setpoint
5.00 ppm
P-Band
0.10 ppm
Reset Time
0 Sec
Derivative Time
0 Sec
Control Timeout
0 Min
Function
Control upw., dnw.
Parameter
pH
Actuator
Time proportional
Control Parameters
5.3.2.32.4.11
Setpoint
7.00 pH
P-Band
0.10 pH
Reset Time
0
Sec Derivative Time 0
Sec Control Timeout 0
Min
Function
Control upw., dnw.
Parameter
Temperature
Actuator
Time proportional
Control Parameters
5.3.2.32.4.12
Setpoint
30.0 ° C
P-Band
1.0 ° C
Reset Time
0
Sec Derivative Time 0
Sec Control Timeout 0
Min
Function
Control upw., dnw.
Parameter
Sample flow
Actuator
Time proportional
Control Parameters
5.3.2.32.4.13
Setpoint
300 rpm
P-Band
30 rpm
Reset Time
0
Sec Derivative Time 0
Sec Control Timeout 0
Min
100
AMI Trides
Operator’s Manual, A-96.250.101
Содержание A-85.119.010
Страница 10: ...1 AMI Trides keys 2 AMI Trides Operator s Manual A 96 250 101...
Страница 15: ...2 Safety Instructions General 2 2 1 AMI Trides Operator s Manual A 96 250 101 7...
Страница 16: ...2 Safety Instructions Warning Notices 2 2 corrosive 8 AMI Trides Operator s Manual A 96 250 101...
Страница 18: ...2 2 3 5 3 5 5 Dimethyldantoin 2 3 6 relay 1 relay 2 2 3 7 ESD 10 AMI Trides Operator s Manual A 96 250 101...
Страница 20: ...3 3 1 2 AMI Trides pH 3 2 3 12 AMI Trides Operator s Manual A 96 250 101...
Страница 22: ...3 Structure and Function Hardware 3 2 AMI Trides LCD PH 14 AMI Trides Operator s Manual A 96 250 101...
Страница 25: ...3 Structure and Function Hardware 6 mm inside dia 14 mm 3 PH pH AMI Trides Operator s Manual A 96 250 101 17...
Страница 32: ...3 Structure and Function Software 24 AMI Trides Operator s Manual A 96 250 101...
Страница 40: ...5 Diagnostics Menu Diagnostics 2 32 AMI Trides Operator s Manual A 96 250 101...
Страница 64: ...6 Maintenance Repair Calibration 56 AMI Trides Operator s Manual A 96 250 101...
Страница 76: ...7 Operation Logger 4 4 optional 68 AMI Trides Operator s Manual A 96 250 101...
Страница 123: ...8 Installation Instrument Setup AMI Trides Operator s Manual A 96 250 101 115...
Страница 128: ...9 Shut Down Transport Disposal Disposal 120 AMI Trides Operator s Manual A 96 250 101...