Upon deciding to directly install the TRIO MP-245A system in your rig, it is useful to follow 

the initial setup procedure to learn how to move the units to allow easy access to the 
mounting screws. 
1.

 

With the power switch on the back of the ROE in the OFF (0) position, connect the power 
adapter’s 24VDC cable to the POWER receptacle. 

 

  

 

Figure 2-3. Rear of TRIO MP-245A ROE/Controller cabinet 

2.

 

With the power OFF (rear panel switch in the “0” position), connect a well-

grounded/earthed wire to the GROUND banana plug receptacle. 

3.

 

With the power OFF, connect the male end of the DB-25 cable to the MANIPULATOR 

connector on the ROE, the other end of which is connected to the TRIO MP-245/M 
micromanipulator electromechanical. (See cautionary note below.) 

4.

 

Verify that the six switches on the rear of the ROE are set as desired. 

5.

 

Power up the system by moving the power switch on the rear of the ROE to the “1” 

position. 

*  

 CAUTION:  NEVER CONNECT OR DISCONNECT THE ROE/CONTROLLER 

FROM THE TRIO MP-245/M WHILE THE POWER IS ON! 

2.5

 

ROE/Controller Rear Panel Controls and Configuration 

2.5.1

 

Power Switch 

The power switch for the TRIO MP-245A system is located on the rear panel of the 
ROE/controller.  At power up, the microprocessor in the ROE/controller scans the attached 
equipment and configures the system accordingly.   

Содержание TRIO MP-245A

Страница 1: ...ice 415 883 0128 Web www sutter com Fax 415 883 0572 Email info sutter com TRIO MP 245A Three Axis Motorized Micromanipulator System With Synthetic Fourth D Axis and USB Interface for External Control...

Страница 2: ...age shows a TRIO MP 245A ROE controller and a TRIO MP 245 M micromanipulator The Sutter Instrument IPA Headstage shown mounted on the micromanipulator is not included in the TRIO MP 245A Series system...

Страница 3: ......

Страница 4: ......

Страница 5: ...ays recommended NOTE There are no user replaceable fuses in the TRIO MP 245A system The TRIO MP 245A system s power supply consists of an external AC to DC switching power adapter If the external powe...

Страница 6: ...art of the micromanipulator electromechanical device while it is in operation and moving Doing so can result in physical injury e g fingers can be caught and pinched between the moving parts of the mi...

Страница 7: ...figuration 15 2 5 1 Power Switch 15 2 5 2 Rear Panel Configuration Switches 16 2 5 2 1 Rear Panel Switches 1 2 3 and 4 Knob Rotation Directionality for Forward Movement 16 2 5 2 2 Rear Panel Switch 5...

Страница 8: ...Command 27 4 7 2 Move to Controller Defined HOME Position h Command 28 4 7 3 Move to Controller Defined WORK Position w Command 28 4 7 4 Move to Specified Home Position H Command 28 4 7 5 Move to Spec...

Страница 9: ...Relative mode 22 Figure 3 10 Relative mode 22 Figure 3 11 Absolute mode 22 Figure 3 12 Angled side view of TRIO MP 245 M to change headstage mount 24 Figure 6 1 Locations of the axis connectors insid...

Страница 10: ...ommand 31 Table 4 14 Move to specified Z axis position z or Z command 31 Table 4 15 Set the angle A command 32 Table 4 16 Recalibrate R command 32 Table 4 16 Straight Line Move S Command Speeds 34 Tab...

Страница 11: ...TRIO MP 245A system Chapter 2 Installation which describes how to install set up and configure all components of the system Chapter 3 Operations which describes how to operate the TRIO MP 245A Chapte...

Страница 12: ...pendent axes X Y and Z each with 25mm travel with a virtual fourth axis D for coaxial pipette movement utilizing a tangent function factoring the holder s angle and the X and Z axes Sub micron 100nm r...

Страница 13: ...mode the display turns red and ROE input is locked out to avoid any accidental motion Pressing and holding the RELATIVE button for three seconds at any location causes the display coordinates to all z...

Страница 14: ...12 TRIO MP 245A FOUR AXIS MICROMANIPULATOR SYSTEM OPERATION MANUAL REV 3 12B 20191001 This page intentionally left blank...

Страница 15: ...adapter plate using four M3 5x6 hex head locking screws Figure 2 1 Side view of TRIO MP 245 M showing mounting adapter plate and lock screws The TRIO MP 245 M is shipped with the adapter plate in plac...

Страница 16: ...fits into the dovetail not shown All the headstage adapters and mounting hardware are included with the manipulator and are shipped in a zip lock plastic bag 2 3 Other Accessories One or more accessor...

Страница 17: ...ug receptacle 3 With the power OFF connect the male end of the DB 25 cable to the MANIPULATOR connector on the ROE the other end of which is connected to the TRIO MP 245 M micromanipulator electromech...

Страница 18: ...ile homing Table 2 2 Rear Panel Configuration Switch 5 Y Axis Homing Movement Lock Out Rear Panel Switch Y Axis Homing Movement Lock Out Enabled Disabled 5 OFF Up ON Down Factory default recommended n...

Страница 19: ...2 Electromechanical device compatibility Internal Switch Definition State Setting Position 2 Eletromechanical device compatibility MP 845 M MP 845S M MP 245 M series OFF UP MP 285 M series micromanipu...

Страница 20: ...18 TRIO MP 245A FOUR AXIS MICROMANIPULATOR SYSTEM OPERATION MANUAL REV 3 12B 20191001 This page intentionally left blank...

Страница 21: ...showing startup screen When starting the TRIO MP 245A system for the first time or if the HOME position has not yet been defined saved the values of all four axes will be 1 000 micrometers microns Fi...

Страница 22: ...dovetail bracket rotate the holder to the desired angle and then retighten the set screw Measure the angle of the holder Tip Many smart phones have an app with a level that can assist the user 3 3 4...

Страница 23: ...the movement will be as follows NOTE Movement to the Home position works only if X coordinates of the HOME position are less than the WORK position HOME and WORK positions cannot be the same 1 Moveme...

Страница 24: ...reen To switch to relative coordinates press the RELATIVE button once To reset the current position to all zeroes depress the RELATIVE for 3 seconds or until a beep is heard and then release the butto...

Страница 25: ...e Movements HOME while moving to Home After Move to Home has been initiated and while the move is in progress pressing HOME a second time pauses the manipulator Pressing HOME again resumes movement 3...

Страница 26: ...loosen the screw in the center of the holding bracket Slide the headstage upward out of the dovetail groove Make any adjustments needed of the headstage and then tighten down but do not over tighten...

Страница 27: ...troller over USB can be conducted 1 USB Direct D2XX mode or 2 Serial RS 232 asynchronous via the VCP device driver VCP mode The first method requires that the VCP device driver not be installed or if...

Страница 28: ...version between microsteps and microns micrometers is the responsibility of the software running on the host computer see Microns microsteps conversion table for conversion factors Microsteps are stor...

Страница 29: ...double precision floating point values Table 4 3 Ranges and bounds Device Axis Millimeters Microns Microsteps MP 845 M MP 845S M or MP 245 M micromanipulator X Y Z 0 25 0 25 000 0 266 667 MP 285 M mic...

Страница 30: ...e to controller defined HOME position h command Tx Delay Rx Ver Total Bytes Byte Offset Len Value Alt key pad Ctrl char ASCII def char Description Dec Hex Binary Tx All 1 0 104 68 0110 1000 0104 h Com...

Страница 31: ...mand 1 4 X steps 5 4 Y steps 9 4 Z steps Rx All 1 0 13 0D 0000 1101 M CR Completion indicator 4 7 6 Move in Straight Line to Specified Position at Specified Speed S Command This command instructs the...

Страница 32: ...110728346 56 25 7 2 5000 2500 0 2500000 0 098425197 50 00 6 2 1875 2187 5 2187500 0 086122047 43 75 5 1 8750 1875 0 1875000 0 073818898 37 50 4 1 5625 1562 5 1562500 0 061515748 31 25 3 1 2500 1250 0...

Страница 33: ...er Total Bytes Byte Offset Len Value Alt key pad Ctrl char ASCII def char Description Dec Hex Binary Tx All 5 0 121 or 91 79 or 5B 0111 1001 or 0101 1011 0121 or 0091 y or Y Command 1 4 Y steps Rx 1 0...

Страница 34: ...sending a command sequence and before sending the next command 3 Clearing Send Receive Buffers Clearing purging the transmit and receive buffers of the I O port immediately before sending any command...

Страница 35: ...LabVIEW always handles byte strings in Big Endian byte order irrespective of operating system and CPU requiring that the four bytes containing a microsteps value be reverse ordered before after conve...

Страница 36: ...8750 1875 0 1875000 0 073818898 37 50 4 1 5625 1562 5 1562500 0 061515748 31 25 3 1 2500 1250 0 1250000 0 049212598 25 00 2 0 9375 0937 5 937500 0 036909449 18 75 1 0 6250 0625 0 625000 0 024606299 12...

Страница 37: ...ect all cables and connections to make sure that all connections are made well and that all connectors are well and evenly seated 6 RECONFIGURATION 6 1 Changing the Rotary Knob Functions on the ROE Co...

Страница 38: ...36 TRIO MP 245A FOUR AXIS MICROMANIPULATOR SYSTEM OPERATION MANUAL REV 3 12B 20191001 This page intentionally left blank...

Страница 39: ...buse misuse or unauthorized repairs will void this warranty Warranty work will be performed only at the factory The cost of shipment both ways is paid for by Sutter Instrument during the first three m...

Страница 40: ...for rod OD 2 4mm MP ROD Rod holder for rod OD 6 25 mm or larger MP RISER 0 5 inch riser MP RISER 1 0 1 inch riser1 MT 78 FS Large fixed stage platform MT 78 FS M6 Large fixed stage platform with M6 ta...

Страница 41: ...ontroller Rear Panel Port Connector Receptacle Cable Connector Types Connects to Cable Type Cable Max Length Power Adapter 3 pin male connector 3 pin power standard female 3 pin male Geographical regi...

Страница 42: ...40 TRIO MP 245A MICROMANIPULATOR SYSTEM QUICK REFERENCE REV 3 12B 20191001 FW V3 12 Weight TRIO MP 245A ROE controller 2 3 lbs 1 04 kg TRIO MP 245 M micromanipulator 3 5 lbs 1 6 kg...

Страница 43: ...nal 1 4 Switch 3 OFF up for MP 845 S M MP 245 M ON down for MP 285 M related devices 3DMS MT 78 MOM SOM objective movers Contact Sutter Instrument tech support for more information Sw Definition State...

Страница 44: ...eted indicator Command Sequence Formatting Each command sequence consists of at least one byte the first of which is the command byte Those commands that have parameters or arguments require a sequenc...

Страница 45: ...e precision floating point values Table D 5 Ranges and bounds Device Axis Len mm Origin Microns Micromet ers m Microsteps steps MP 845 M MP 845S M or MP 245 M micromanipulator X Y Z 25 BOT 0 25 000 0...

Страница 46: ...s 5 4 Y steps 9 4 Z steps Rx All 1 0 13 0D 0000 1101 M CR Completion indicator Move in Straight Line to Specified Position at Specified Speed S Tx All 14 0 83 53 1001 0111 0083 S Move all three axes s...

Страница 47: ...lute vs Relative Positioning Current position c and move commands always use absolute positions All positions can be considered relative to the Origin Position 0 but all are in fact absolute positions...

Страница 48: ...5000000 0 196850394 100 00 14 4 6875 4687 5 4687500 0 184547244 93 75 13 4 3750 4375 0 4375000 0 172244094 87 50 12 4 0625 4062 5 4062500 0 159940945 81 25 11 3 7500 3750 0 3750000 0 147637795 75 00 1...

Страница 49: ...mplete indicator 34 external control intercommand delay 34 external control clearing send receive buffers 35 external control positions in microsteps 35 external control ranges and bounds 35 external...

Страница 50: ...Switch 1 Reserved 17 Switch 3 Reserved 18 Switch 4 Linear nonlinear manual operation 18 internal configuration Switch 2 electromechanical type 18 internal configuration switches 17 linear nonlinear m...

Страница 51: ...ging rotary knob functions on the ROE 39 Reserved configuration switches internal 1 3 17 18 S safety warnings mains fuse 3 safety warnings precautions operational 4 SAFETY WARNINGS PRECAUTIONS 3 elect...

Страница 52: ...50 TRIO MP 245A MICROMANIPULATOR SYSTEM QUICK REFERENCE REV 3 12B 20191001 FW V3 12 NOTES...

Отзывы: