TRIO MP-245A FOUR-AXIS MICROMANIPULATOR SYSTEM OPERATION MANUAL – REV. 3.12B (20191001)
15
Upon deciding to directly install the TRIO MP-245A system in your rig, it is useful to follow
the initial setup procedure to learn how to move the units to allow easy access to the
mounting screws.
1.
With the power switch on the back of the ROE in the OFF (0) position, connect the power
adapter’s 24VDC cable to the POWER receptacle.
Figure 2-3. Rear of TRIO MP-245A ROE/Controller cabinet
2.
With the power OFF (rear panel switch in the “0” position), connect a well-
grounded/earthed wire to the GROUND banana plug receptacle.
3.
With the power OFF, connect the male end of the DB-25 cable to the MANIPULATOR
connector on the ROE, the other end of which is connected to the TRIO MP-245/M
micromanipulator electromechanical. (See cautionary note below.)
4.
Verify that the six switches on the rear of the ROE are set as desired.
5.
Power up the system by moving the power switch on the rear of the ROE to the “1”
position.
*
CAUTION: NEVER CONNECT OR DISCONNECT THE ROE/CONTROLLER
FROM THE TRIO MP-245/M WHILE THE POWER IS ON!
2.5
ROE/Controller Rear Panel Controls and Configuration
2.5.1
Power Switch
The power switch for the TRIO MP-245A system is located on the rear panel of the
ROE/controller. At power up, the microprocessor in the ROE/controller scans the attached
equipment and configures the system accordingly.
Содержание TRIO MP-245A
Страница 3: ......
Страница 4: ......
Страница 52: ...50 TRIO MP 245A MICROMANIPULATOR SYSTEM QUICK REFERENCE REV 3 12B 20191001 FW V3 12 NOTES...