TRIO MP-245A FOUR-AXIS MICROMANIPULATOR SYSTEM OPERATION MANUAL – REV. 3.12B (20191001)
17
2.5.1
Internal Configuration Switches
Figure 2-5. Internal configuration switches (switch positions shown are factory defaults).
2.5.1.1
Internal Switch 1 Reserved
Internal Switch 1 is unused and reserved for future use.
2.5.1.2
Internal Switch 2: Electromechanical Device Compatibility
.
Table 2-4. Internal Configuration Switch 2: Electromechanical device compatibility.
Internal
Switch #
Definition
State
Setting Position
2
Eletromechanical
device
compatibility
MP-845/M, MP-845S/M, & MP-245/M series
OFF*
UP*
MP-285/M series micromanipulator; 3DMS or MT-78
stage; MOM & SOM objective mover)
ON
DOWN
* Factory default (recommended normal operation setting)
2.5.1.3
Internal Switch 3 Reserved
Internal Switch 3 is unused and reserved for future use.
2.5.1.4
Internal Switch 4 Linear/Nonlinear Manual Operation
Internal Switch 4 switches between linear and nonlinear manual operation. Contact Sutter
Instrument technical support for more information.
1 2 3 4