background image

 

16 

3.2.2

 

LED Control Switch 

This is a three-position toggle switch that is used to manually turn the LED on, off, or 
activate TTL switching through the TTL-IN BNC connector. 
ON: Manually turns on the LED for manual control.  
OFF: The intensity of the LED is controlled via voltage range of 0-5 VDC applied to 
ANALOG-IN.  
TTL: LED on/off state is controlled by TTL triggering (via TTL-IN). Note that remote analog 
control (via ANALOG-IN) is turned off when using TTL-IN. 

3.2.3

 

ANALOG-IN Control Switch 

Selects between LED dimming control through the ANALOG-IN BNC connector (REMOTE) 
or manually using the INTENSITY (PWM) control knob (LOC-1). 
REMOTE: Allows for the dimmed state of the LED to be controlled by a voltage (0 – 5 VDC) 
applied to ANALOG-IN. The LED Control Switch must be set to OFF in order for dimming 
via analog remote input to function properly. When controlling the LED’s dimmed state in 
this mode, the status display turns blue. 
LOC-1: Allows the INTENSITY (PWM) knob to control the dimmed state of the LED. For the 
INTENSITY (PWM) knob to be effective, the LED Control Switch must be set to ON or TTL. 

3.2.4

 

INTENSITY (PWM) Control Knob 

Intensity (PWM) Knob: This knob dims the LED via Pulse Width Modulation. In essence, the 
LED is being switched at 30KHz. This is the preferable mode of dimming for most 
applications, as the LED will run cooler, and exhibit no color shift in the special output as the 
light is dimmed. For the INTENSITY (PWM) knob to be effective, the LED Control Switch 
must be set to ON or TTL. 
The control of the intensity via PWM can also be controlled externally by connecting a 0-5 
VDC source to ANALOG-IN connector. This external control is enabled by turning the 
ANALOG-IN switch to the LOC-1 position and the LED switch to either ON or TTL. Note 
that when PWM intensity control is set up for external control, the INTENSITY (PWM) 
knob is disabled. 

3.2.5

 

CURRENT Control Knob 

Current Knob: Dims LED by adjusting the current delivered to the LED. Current control 
may cause color shift in the LED’s spectral output. However, current-controlled dimming 
may be preferable in situations involving the use of ultra high-speed cameras. For the 
CURRENT knob to be effective, the LED Control Switch must be set to ON or TTL.  
Note that the CURRENT knob is functional regardless of the positions of the LED and 
ANALOG-IN switches, and manual or external state of intensity via PWM (manually 
adjusting INTENSITY (PWM) knob or applying 0-5 VDC on ANALOG-IN input). 

LAMBDA HPX-L5 OPERATION MANUAL – REV. 1.03  (20161130) 

Содержание Lambda HPX-L5

Страница 1: ...igh Power LED Light Source System Air Cooled with Liquid Light Guide Operation Manual Rev 1 03 20161130 One Digital Drive Novato CA 94949 Voice 415 883 0128 Web www sutter com Fax 415 883 0572 Email i...

Страница 2: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 2 Copyright 2016 Sutter Instrument Company All Rights Reserved...

Страница 3: ......

Страница 4: ......

Страница 5: ...uppression is recommended at all times Fuse Replacement Replace only with the same type and rating 2 Amp 250V 5 x 20mm Time Delay fuse EIC 60127 2 Example Bussmann GDC 2A or S506 2A RoHS A spare fuse...

Страница 6: ...re is any obvious damage to any part of the instrument Avoiding Physical Injury while Powered up and Emitting Light WARNING DO NOT LOOK DIRECTLY INTO THE OUTPUT OF THE LIGHT APERTURE OF THE HIGH POWER...

Страница 7: ...ED On Off State while Transitioning Between LED Switch Positions ON and TTL Switch 3 12 2 3 4 Firmware and Lambda HPX Model Switch 4 13 3 OPERATIONS 15 3 1 Indicators 15 3 1 1 Status Display 15 3 2 Co...

Страница 8: ...Liquid light guide installation 10 Figure 2 2 Lambda HPX series left side panel removed for access to configuration DIP switches 11 Figure 2 3 Configuration DIP switch bank as viewed head on and top d...

Страница 9: ...onnectors The Lambda HPX L5 is liquid light guide coupled version of the Lambda HPX high output LED light source illuminator The liquid light guide allows for LED cooling via an internal heatsink with...

Страница 10: ...he MPX L5 Integral shuttering time is 10 microseconds to turn on and off TTL input and output allows for triggering from either software or directly from another device such as a camera or digital I O...

Страница 11: ...s a black end sleeve preinstalled over the brass ferrule on each end The two ends of the light guide are identical The end sleeves are designed to slide into the mounting flanges on the input and outp...

Страница 12: ...couple the light into the microscope In external adapters the adapter system mounts in place of the manufacturer s lamp house typically on the back of the microscope In internal adapters the adapter...

Страница 13: ...ing the first three switches of the four switch DIP bank located on the control board inside the controller cabinet To access the DIP switch bank remove left side panel left side while viewing the fro...

Страница 14: ...Switch 2 determines how the state at TTL IN determines the on off state of the lamp when the LED Control Switch is in the TTL position When ON TTL IN going HIGH turns the LED on TTL IN going LOW turn...

Страница 15: ...p on ON Lamp turns on Notes 1 and 2 3 Lamp state when switching from manual control to TTL OFF Lamp turns off Note 1 ON Lambda HPX L5 4 Lambda HPX model specific operation Note 3 OFF Lambda HPX Defaul...

Страница 16: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 14 This page intentionally blank...

Страница 17: ...l indicators controls and connectors on the Lambda HPX L5 controller 3 1 Indicators 3 1 1 Status Display Displays intensity as a percentage and current level in terms of amperes Intensity 5 Current 18...

Страница 18: ...s knob dims the LED via Pulse Width Modulation In essence the LED is being switched at 30KHz This is the preferable mode of dimming for most applications as the LED will run cooler and exhibit no colo...

Страница 19: ...s a high signal the LED turns on for the duration that TTL IN is held high 3 3 4 TTL OUT BNC Connector Outputs a triggering signal for external software When the LED is off TTL OUT is high when the LE...

Страница 20: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 18 This page intentionally blank...

Страница 21: ...L5 system is required to prevent excessive dust accumulations Wipe all exterior surfaces with a dry soft cotton cloth Periodically inspect all cables ensuring that all connections are made well and co...

Страница 22: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 20 This page intentionally blank...

Страница 23: ...only at the factory The cost of shipment both ways is paid for by Sutter Instrument during the first three months this warranty is in effect after which the cost is the responsibility of the customer...

Страница 24: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 22 This page intentionally blank...

Страница 25: ...utter Information for wavelength availability and other information 4 4 Microscope Adapters Liquid light guide adaptors for various microscopes are available Please contact Sutter Instrument for more...

Страница 26: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 24 This page intentionally blank...

Страница 27: ...above the fuse holder and below the power receptacle and pry the holder straight out of the power entry module Figure 4 1 Power entry module and fuse location on the rear panel of the Lambda HPX L5 c...

Страница 28: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 26 This page intentionally blank...

Страница 29: ...put 400 700 nm white light Lamp life 25 000 hours at 80 original output Input Output TTL and analog Dimming PWM or Current Control LED Wavelength Cool white Optional LEDs 630 530 460 and 405 nm Electr...

Страница 30: ...0mm Time Delay fuse EIC 60127 2 E g Bussmann GDC S506 2 R or S506 2A RoHS Other Dimensions Controller 12 25 x 9 x 5 25 in 311 x 229 x 133 mm Head attaches to scope 2 5 x 3 in 64 x 76 x 127 mm Length o...

Страница 31: ...ains 28 fuses replacement mains 3 I indicators 15 status display 15 input analog 27 TTL 27 voltage 28 installation 9 unpacking 9 Installation 9 instructions 9 introduction 7 description 7 features 7 t...

Страница 32: ...entry module 25 power switch 16 precautions 3 electrical 3 PWM frequency 27 S safety warnings 3 electrical 3 safety warnings precautions operational 4 safety warnings precautions 3 4 shuttering 27 st...

Отзывы: